The 1.5 ns fwhm pulses delivered with 2.5 MHz repetition rate by the CN Van-de-Graaff accelerator of CBNM have been further compressed. The time compression is based on a two-gap buncher powered with a frequency of 105 MHz. The synchronization with a pickup electrode signal is realized by a gate phase lock device which is described as well as the electronics used to perform time resolution measurements. Pulses with fwhm <or=0.2 ns were obtained.

Additional information

Authors: CRAMETZ A, JRC GEEL ESTAB. (BELGIUM);MAX-PLANCK-INSTITUT FUER KERNPHYSIK, HEIDELBERG (GERMANY);DE JONGE S, JRC GEEL ESTAB. (BELGIUM);MAX-PLANCK-INSTITUT FUER KERNPHYSIK, HEIDELBERG (GERMANY);LISKIEN H, JRC GEEL ESTAB. (BELGIUM);MAX-PLANCK-INSTITUT FUER KERNPHYSIK, HEIDELBERG (GERMANY);JAESCHKE E, JRC GEEL ESTAB. (BELGIUM);MAX-PLANCK-INSTITUT FUER KERNPHYSIK, HEIDELBERG (GERMANY);INGWERSEN H JRC GEEL ESTAB. (BELGIUM), JRC GEEL ESTAB. (BELGIUM);MAX-PLANCK-INSTITUT FUER KERNPHYSIK, HEIDELBERG (GERMANY);MAX-PLANCK-INSTITUT FUER KERNPHYSIK, HEIDELBERG (GERMANY), JRC GEEL ESTAB. (BELGIUM);MAX-PLANCK-INSTITUT FUER KERNPHYSIK, HEIDELBERG (GERMANY)
Bibliographic Reference: NUCLEAR INSTRUMENTS AND METHODS IN PHYSICS RESEARCH, A VOL. 242 (1986), PP. 179-184
Record Number: 1989124088700 / Last updated on: 1987-01-01