Community Research and Development Information Service - CORDIS

Abstract

The report summarizes the results of six laboratories who calibrated accelerometers from three different makers. The range of frequency covered extended from 10 Hz to 10 kHz. The results of these four laboratories who used laser interferometry agreed to within 0.5% in the range of frequency from 40 to 6300 Hz. One of the three accelerometers although produced to serve as a transfer standard was the cause of difficulties: - instability: there was a drift of 1% at some particular period during circulation; - errors of 0.5% can be made because of its sensitivity to mechanical mounting conditions.

Additional information

Authors: CANTON H CEA, BELIN-BELIET, PARIS (FRANCE) BAGGIA S ISTITUTO DI METROLOGIA "G.COLONNETTI", TORINO (ITALY) BRINKLEY J AERONAUTICAL QUALITY ASSURANCE DIRECTORATE, HAREFIELD (UK) MARCHANDISE H CEC BRUXELLES (BELGIUM) , CEA, BELIN-BELIET, PARIS (FRANCE);ISTITUTO DI METROLOGIA "G.COLONNETTI", TORINO (ITALY);AERONAUTICAL QUALITY ASSURANCE DIRECTORATE, HAREFIELD (UK);CEC BRUXELLES (BELGIUM)
Bibliographic Reference: EUR 10691 EN (1986) MF, 74 P., BFR 150, BLOW-UP COPY BFR 375, EUROFFICE, LUXEMBOURG, POB 1003
Availability: Can be ordered online
Record Number: 1989124133600 / Last updated on: 1987-01-01
Category: PUBLICATION