Abstract

This report summarizes the results of six laboratories of the Community which calibrated with the highest possible accuracy resistors of nominal values of 1 G, 100 G, 10 T and 100 T. The measurements were done at 10 V, 100 V and 1,000 V. These measurements are difficult, specially at higher voltages. When measurements are done under 10 V, the uncertainty of the measurements ranges from 10-5 for 10 G to 10-3 at 100 T, although some laboratories had higher uncertainties. The discrepancies between laboratories tend to increase with the voltage.