Any mass spectrometric measurement system has inherent inaccuracies caused by isotope fractionation in the ion source and non-linearity of the detector and or measurement system. A set of well known triple isotope mixtures is an excellent means to verify and correct non- linearity of the mass spectrometer detector and measurement system after the isotope fractionation has been corrected for. CBNM IRM 072/1-15 is a set of such mixtures. Its preparation and characterization are described.

Additional information

Authors: LYCKE W, CENTRAL BUREAU FOR NUCLEAR MEASUREMENTS, GEEL (BELGIUM);DE BIEVRE P, CENTRAL BUREAU FOR NUCLEAR MEASUREMENTS, GEEL (BELGIUM);VERBRUGGEN A, CENTRAL BUREAU FOR NUCLEAR MEASUREMENTS, GEEL (BELGIUM);HENDRICKX F, CENTRAL BUREAU FOR NUCLEAR MEASUREMENTS, GEEL (BELGIUM);ROSMAN K CENTRAL BUREAU FOR NUCLEAR MEASUREMENTS, GEEL (BELGIUM), CENTRAL BUREAU FOR NUCLEAR MEASUREMENTS, GEEL (BELGIUM)