Abstract

The report describes laboratory comparison of calibrations of 400 mm and 900 mm gauge blocks involving six EEC national metrology laboratories. After correcting all measurements to 20 C, good agreement was found between the laboratories for both the length and centre line positions of both gauge blocks. This is a collection of the individual reports of the laboratories which participated in the comparison campaign. A synthesis report is published full size with the same title.

Additional information

Authors: FISCHER W, Physikalisch-Technische Bundesanstalt, Braunschweig (DE);BIRCH K G, National Physical Laboratory, Teddington (GB);GILL P, National Physical Laboratory, Teddington (GB);CANTA G, Istituto di Metrologia "G. Colonnetti", Torino (IT);MOREI S, Istituto di Metrologia "G. Colonnetti", Torino (IT);PRESTINICE A, Istituto di Metrologia "G. Colonnetti", Torino (IT);VOORHOF H, Ministère des Affaires Economiques, Service de la Métrologie, Bruxelles (BE)
Bibliographic Reference: EUR 13770 EN (1991) 11 pp., MF, ECU 4
Availability: (2)
Record Number: 199110937 / Last updated on: 1994-12-02
Category: PUBLICATION