Community Research and Development Information Service - CORDIS

Abstract

A European intercomparison of ultrasonic power measurements is being carried out. Stage 1 involving a 4 MHz transducer is reported here. The laboratories participating in this project are the National Physical Laboratory (NPL, United Kingdom), the Medical Technology Unit of TNO (The Netherlands) and the Physikalisch-Technische Bundesanstalt (PTB, Germany). The quantity of interest is the electroacoustic radiation conductance G of the transducer when radiating into water at room temperature. The final G result, averaged from the measurements of all participants, is 45.64 microS. The deviations from this mean value are -3.4% for NPL, +0.2% for TNO and +3.2% for PTB. Possible reasons for these differences are discussed and proposals for future action are made. The 4 MHz transducer is found to be well-suited for the purposes of a power standard. There have been no indications of a temporal change since the transducer construction in 1984, and the radiation conductance is independent of the signal amplitude within the limits of repeatability. The published synthesis report is EUR 12681.

Additional information

Authors: OOSTERBAAN W A, TNO Medical-Technology Unit, Leiden (NL);SHAW A, National Physical Laboratory, Teddington (GB);SMITH R A, National Physical Laboratory, Teddington (GB)
Bibliographic Reference: EUR 13860 EN (1991) 49 pp., MF, ECU 4
Availability: (2)
Record Number: 199111379 / Last updated on: 1994-12-02
Category: PUBLICATION
Original language: en
Available languages: en