Community Research and Development Information Service - CORDIS

  • European Commission
  • CORDIS
  • Publications
  • The certification of the catalytic concentration of alanine aminotransferase (ALT) (E.C. 2.6.1.2) from pig heart in a reconstituted lyophilized material CRM 426

Abstract

The report describes the work carried out for certification of the catalytic concentration of alanine aminotransferase (L-alanine: 2-oxoglutarate aminotransferase, EC 2.6.1.2) (ALT) in a lyophilised solution of partially purified ALT from pig heart dissolved in a bovine serum albumin matrix. The preparation has been shown to be homogeneous, and the loss of activity on storage at -20 C is predicted to be less than 0.01% per annum on the basis of accelerated degradation studies. The catalytic concentration of ALT in samples of the reconstituted material is certified to be 129 U/l or 2.14 microkat/l reconstituted solution according to the IFCC recommended method at 30 C.

Additional information

Authors: SCHIELE F, Centre du Médecine Préventive, Vandoeuvre-les-Nancy (FR);SIEST G, Centre du Médicament, Nancy (FR);COLINET E, CEC Bruxelles (BE);PROFILIS C, CEC Bruxelles (BE)
Bibliographic Reference: EUR 14475 EN (1992) 52 pp., MF, ECU 4, blow-up copy ECU 9
Availability: (2)
Record Number: 199211248 / Last updated on: 1994-11-29
Category: PUBLICATION
Original language: en
Available languages: en