Abstract

Instructions are provided for a new reference material, SCA91 of Cu and Au, to calibrate AES instrument intensity scales in the differential mode. The calibration measurements are made at 10 eV peak-to-peak modulation using a 1 to 5 microamp, 5 keV beam set at 30 to the sample normal. The operating resolution of the spectrometer must be measured and this is completed with 5 eV and 10 eV measures of the Cu 916 eV peak.

Additional information

Authors: SEAH M P, National Physical Laboratory, Teddington (GB)
Bibliographic Reference: EUR 15422 EN (1994) 17 pp., MF, ECU 4
Availability: (2)
Record Number: 199510288 / Last updated on: 1995-03-08
Category: PUBLICATION
Original language: en
Available languages: en