Community Research and Development Information Service - CORDIS

Abstract

A procedure for the preparation of a fast response Ag/Ag(2)SO(4) reference electrode to be used for the detection of the endpoint of a coulometric titration is described. The reference electrode consists of a platinum wire covered with a film of silver, obtained by thermal decomposition of silver oxide which is then coated with silver sulphate by electrolytic oxidation in a solution of H(2)SO(4).

Additional information

Authors: FINOLY G, JRC Geel (BE);LEIDERT W, JRC Geel (BE);BICKEL M, JRC Geel (BE)
Bibliographic Reference: Article: Fresenius' Journal of Analytical Chemistry (1995)
Record Number: 199511300 / Last updated on: 1995-11-23
Category: PUBLICATION
Original language: en
Available languages: en