Abstract

The operation of the future on-site laboratory (OSL) at Sellafield will have to rely on quality control measures and all plutonium oxide samples will be measured by a neutron-gamma counter (OSL-counter) which combines the technique of passive neutron-coincidence counting and high-resolution gamma spectrometry.

A batch of plutonium oxide with isotopic characteristics close to those expected in material in the OSL was prepared and certified, and used for calibrating the OSL-counter. The paper presents the verification and the evaluation of the precision and the accuracy of the different analytical techniques applied for the plutonium element assay and isotopic abundance determinations. .

Additional information

Authors: CROMBOOM O, JRC Karlsruhe (DE);KOCH L, JRC Karlsruhe (DE);VAN GEEL J, JRC Karlsruhe (DE);MAYER K, JRC Geel (BE);DE BIEVRE P, JRC Geel (BE);OTTMAR H, Forschungszentrum KarlsruheGmbH (DE)
Bibliographic Reference: Paper presented: 5th International Conference on Facility Operations-Safeguards Interface, Jackson Hole (US), September 24-29, 1995
Availability: Available from (1) as Paper EN 39175 ORA
Record Number: 199512043 / Last updated on: 1996-01-15