Community Research and Development Information Service - CORDIS

Abstract

A current interest in chemistry concerns traceability of analytical measurements to the International System of Units (SI) and the estimation of their uncertainties in accordance with principles of metrology, that is measurement science. Primary methods of measurement achieve traceability to SI directly without intermediate reference standards or materials and without significant empirical correction factors. Isotope-dilution mass spectrometry is such a primary method with the potential of smallest presently achievable uncertainties for many analytical measurements. A simple explanation of the method is given in terms of current concepts, and general equations are derived. The required uncertainly estimation is emphasized in terms of these equations.

Additional information

Authors: DE BIÈVRE P, JRC Geel (BE);PEISER H S, NIST, Gaithersburg (US)
Bibliographic Reference: Article : Analytical Chemistry (1996)
Record Number: 199610916 / Last updated on: 1996-09-16
Category: PUBLICATION
Original language: en
Available languages: en
Follow us on: RSS Facebook Twitter YouTube Managed by the EU Publications Office Top