Abstract

This paper reports the reliability of efficiency calibrations for gamma-ray detectors using the calculation codes ANGLE and LabSOCS. For experimental verification, three HPGe detectors under various laboratory geometry configurations were used for this study. An overall comparison between experimental and calculated efficiency calibration curves is presented and comments on the various error sources affecting the final results are given. The deviations are generally below 10%, which could be acceptable for many applications.