{"contenttype":"result","language":"en","availableLanguages":"en","rcn":"247156","id":"285203_PS","title":"Final Report Summary - SNIFFER (A bio-mimicry enabled artificial sniffer)","teaser":"Dogs are known for their incredible ability to detect odours. This makes them precious assets for a broad variety of security applications. However, dogs can only be trained to a limited set of applications, get tired quickly and for many border control applications are poorly...","description":"final","summary":"Executive Summary:\nDogs are known for their incredible ability to detect odours. This makes them precious assets for a broad variety of security applications. However, dogs can only be trained to a limited set of applications, get tired quickly and for many border control applications are poorly accepted by the public. This has created an increasing drive to develop artificial sniffers able to replace or complement dogs and cover a broader range of applications.\nMany technological approaches have been tried out but without a resounding success to-date. This can be explained by the complexity of the problem at hand:\n\u2022\tSensitivity \/ Selectivity: the need to extract a few molecules from a larger set of molecules\n\u2022\tEnvironmental complexity: the often challenging \u201codorant\u201d background\n\u2022\tMoving targets: the challenge of detecting odours from moving targets\n\u2022\tPracticality: the cost and lack of portability of the technologies employed\n\u2022\tUsability: the often need for the intervention of especially trained staff for the use of the technologies employed\nCurrent approaches to \u0022artificial sniffing\u0022 are constrained in terms of applicability and require extensive training for each new odour. Therefore most of the current efforts aim at integrating complementary technologies, which makes it difficult to propose upgradable inexpensive devices. \nThe SNIFFER project proposes a very different concept offering a comprehensive one-stop shop solution to this need. The key idea is to couple selective binding elements from biological systems to highly sensitive diamond sensor technologies. SNIFFER addresses a broad range of border security challenges by delivering an innovative set of highly effective, flexible, low-cost and portable solutions, integrating, adapting and improving advanced sensor technologies, thereby leveraging and complementing the capabilities of dogs. During its 40 months, the SNIFFER project has achieved the following objectives:\n\n1.\tProvide a range of practical border security solutions to complement dogs and leverage their capabilities. SNIFFER has focused on enabling solutions for specific usage contexts:\n\u2022\tBody-scanning people at airports and at border crossing points\n\u2022\tScanning open luggage\n\u2022\tScanning the interior of cars for illegal substances and explosives\n\u2022\tScanning containers at harbours \n\u2022\tScanning closed luggage on a luggage belt \nThese examples illustrate the breadth of the SNIFFER approach. Several of these solutions were prototypes and evaluated in laboratory tests and in small field trials.\n\n2.\tDevelop artificial sniffer technologies. SNIFFER has delivered significant progress for a range of technologies and corresponding SNIFFER sub-systems as building blocks for future solutions:\n\u2022\tCollection \u0026 Sampling: a new generation of sampling techniques makes it possible to sniff not only gas molecules but also particles\n\u2022\tPreconcentration: micro-preconcentration technologies based on porous silicon micro-channels have been improved and especially adapted to SNIFFER sensing\n\u2022\tLBP - engineering: tailored odorant binding proteins (OBP) bind specific complementary families of molecules\n\u2022\tBiosensors: sensors attained sensitivities in the 1 ppb range by mimicking olfactory mechanisms from nature; they immobilise the OBPs on the current generation of diamond sensors \n\u2022\tLearning systems: multi-parametric training methods are applied to arrays of LBP-enriched diamond sensors support to match macro-patterns against a previously constituted database; this database has been built by exposing the sensor array to a large sample set \n\u2022\tSelf-diagnostic: advanced self-diagnostic features allow for the detection of \u201csensor fatigue\u201d when the sensor has been contaminated and is no longer sensitive enough\nThese technologies have a broad application potential in many areas such as food safety, environment, industrial control and health.\n\n3.\tOptimise efficiency and impact of SNIFFER:\nThe project has actively carried out dissemination activities to engage with third parties. It built a community of users and experts interested in SNIFFER solutions in the European border security community, in particular by interacting regularly with FRONTEX.\nThe project has prepared future exploitation not only by achieving its technological objectives but by coordinated exploitation planning and IPR management.\nThe SNIFFER RTD roadmap worked in 2 iterations to produce 2 sets of devices to support a practical dialogue with the users and solve the integration problems early in the project life:\n\u2022\tA first generation set \u0022V1\u0022 of SNIFFER devices for a first assessment based on concrete metrics provided by the SNIFFER users (including advice from the UAG members);\n\u2022\tA second generation \u0022V2\u0022 built from the assessment of V1 at project end. V2 integrated even more powerful technologies illustrating the full potential of the SNIFFER approach from the user perspective on an additional set of usage cases.\n\nAt project end, SNIFFER Partner CEA won the Wolfgang Gopel Memorial award at the 16th International symposium ISOEN on Olfaction and Electronic noses held in France early July 2015. ISOEN is the only global event dedicated to E-senses (E-nose, E-tongue and E-eye) and to odor generation\/transmission. It awarded Emmanuel Scorsone from CEA, SNIFFER Project Coordinator, with the Wolfgang Gopel Memorial award for: \u201cDiamond micro-cantilevers as transducers for olfaction receptors-based biosensors: application to the receptors M71, OR7D4 and OR1740\u201d presentation.\n\nProject Context and Objectives:\nWork performed since the beginning of the project \nWP1 \u201cSecurity Usage Cases, Metrics \u0026 Validation, Societal \u0026 Ethical Aspects\u201d delivered to the SNIFFER Consortium solid and reliable user requirements and functional specifications of the envisioned devices. WP1 was originally structured by six concrete usage cases. These requirements were validated after a dedicated meeting with the SNIFFER User Advisory Group (UAG) in Manchester in July 2012. During the following 12 months, WP1 continued to provide guidance and feedback to the technical partners on usage issues for the future SNIFFER solutions. \nBetween November 2012 and January 2013 WP1 examined the details and set-up of the midterm field trials and their possible implications. A field trials preparation plan was prepared with valuable contributions from the user partners. Work then started to identify the necessary clearances and authorisations for these trials. The field trial description presented the Lab Testing protocols and the Laboratory Assessment Methodology proposed by ST(SI)2 and CREL. During the Munich meeting in spring 2013, lab trials were also planned to ensure the prototype would be functional and safe to be used in real life conditions.\nAfter the project midterm review the consortium decided to limit the use cases to 3, letting aside Use Case 5 (UC5) which involved containers and human odours detection. The project scope focused on the rest of the use cases in order to develop 2 devices \u2013 instead of 4 originally anticipated \u2013 and to broaden the range of substances to detect.\nFurthermore, instead of running full-scale field trials in the mid-term of the project, a task that required overcoming a number of logistics and authorisations problems, the device was extensively tested in a specialised lab and plan a \u201cclose to real life conditions\u201d field test at the last months of the project.\nPartner 3DSA, WP leader, consolidated the report on National Ethical Approvals that covered the EU privacy and Data Protection Legal Framework and presented the outline for the Trials Procedure Protocols. Initially, extensive work took place to investigate the different ethical and societal issues regarding pursuing security through the detection of odours which became irrelevant once the scope of the project refocused. Furthermore, recommendations were provided in order to ensure the compliance of the SNIFFER devices with ethical and fundamental rights standards by technological design. 3DSA  also investigated the required actions and necessary authorisations for the implementation of the final field trials according to the trial protocols that would be developed. To simplify things and avoid time consuming authorisations, the field trials were restricted to the members of the consortium and a few stakeholders (Greek Police, Athens International Airport auhtorities). These field trials took place in Athens International Airport between 10-12th of March and thanks to the involvement and support from both the AIA\u2019s authorities and the Greek Police, were very successful.\n\t\nWP2 \u201cSNIFFER Device Integration\u201d integrated, delivered and supported operational SNIFFER sensor system prototypes.\nThe integration work package proceeded iteratively with two cycles of specification, integration and testing. Early into the project it delivered refined specifications of functionalities and system design architecture for SNIFFER devices 1 and 2 focusing on body scanning devices and searching technology at open luggage and cars. \nAt project midterm system design in sampling, preconcentration and detection device developments for the V1 version was completed. All related and relevant mechanical and electronic sub-elements had been constructed, integrated and tested. Interface definition, electrical and flow design between sampling, pre-concentrator and detector devices were aligned with all involved system partners. Integration of the sub element units were already going to start and field test design and planning were on a mature level. Major system integration aspects between partners had been gathered. The functionalities of the different sub-systems were aligned and the systematic interaction to achieve the device integration were ensured. All sampling and detection submodules, including the Automatic Trace Collector module, the ESP sampler modules, the pre-concentrator and the main SNIFFER detector module, were completed by the various partners involved and could be integrated altogether successfully. A slight delay in the delivery of the Automatic Trace collector module was observed but the device was finally delivered and this delay had no impact on final integration. Integration was led by CEA with a strong involvement with many partners, and in particular UNIMAN, AIRBUS, ARMINES\/EPFL, and TTS.  \nBased on the experience built from the first version V1 testing, WP2 provided a more detailed specification and description of the devices for the V2 developments, in particular as related to container inspection and luggage belts investigations in airports. This work was carried out in conjunction with the work on use cases in WP1 and resulted in an updated list of devices to be built within the course of the project. Following the feedbacks received from laboratory trials of V1 and end-users observations and comments, a different integration approach to V1 was undertaken in V2. Here the choice was made to sample narcotics or explosives particles using either the ESP devices or the Automatic Trace collector module onto commercially available filters normally used with IMS machines. The trapped particles were then vaporised off the filter using a thermal desorber embedded directly into the SNIFFER detector submodule. Hence a new dedicated desorber was developed by AIRBUS and integrated during the second term of the project. The general integration scheme between the pre-concentrator and the detector device remained unchanged between the V1 and V2.The HMI was developed successfully in parallel to integration work and was first assessed in laboratory conditions and then further validated in field operation at Athens International Airport. \n\nWithin WP3 \u201cSub-system for sampling, pre-concentration and pre-treatment of target analytes\u201d, starting from the key functional requirements for the SNIFFER devices and the physico-chemical properties of the SNIFFER target analytes, the overall functionality and the architecture of the SNIFFER sampling and pre-treatment system was developed at the end of the first project period. \nThe focus of the second half of the project was the new set of enhanced sampling technologies which proved highly relevant towards the end of the project. With the new desorbing design, collected traces of particulates matter now can fast and effectively desorbe into the final OBP detection system. New fluidic components were included based on the pre-SNIFFER experiences with regards to high stickiness of explosive and drug residuals on the wall of sensors gas inlet. Heated glass based sensing inlet and desorbing outlet tubes were used to overcome these issues. Sampling material of the same dimensions and characteristics in addition facilitated matching measurements against different standard airport trace detection systems (e.g. Ionscan 400). \nSustaining the airports and car detection operational requirements a first (ESP) sampling development for version 2 devices was further functionalised towards a more hand-held based sampling system, with battery driven functionalities and COTS based light weight construction figures. The two WP3 device groups developed during the second period of the project, together with the odour binding detection system, demonstrated their performance at Athens airport very successfully.\n\nWP4 \u201cSensor Sub-System \u0026 Array Engineering\u201d initially included the development of a first version of the sensor subsystemand the following activities:\n\u2022\tThe hardware integration of the biosensors\n\u2022\tThe multi-parametric training of the biosensors to optimise the recognition potential from the sensor array\n\u2022\tThe integration of software and hardware into the Sensor Sub-System.\nThese developments of the VA sensor subsystem were achieved at project midterm and during the second period of the project, a version V2 of the sensor sub-system was built within WP4. In this second version, minor changes were made respect to V1, to address some issues that arose during V1 characterisation. These issues compromised the success of the project by reducing considerably the limit of detection of the sensor system. The major changes carried out were:\n\u2022\tThe sample inlet path was shorten by removing the pre-concentrator from the system;\n\u2022\tGlass tube was used for the inlet path instead of stainless steel;\n\u2022\tAn heater and a temperature controller were added to the inlet line to keep it at 100 \u00b0C  \n\u2022\tUSB communication protocol was develop to implement a user interface based on the tablet technology.\nOther modifications were carried out within the directly related WP7 that contributed to improve the sensor sub-system performance:\n\u2022\tThe modularity of the sensor chamber was increased;\n\u2022\tCorruption of SPI bus data between the sensor board and main board was solved by reducing the communication baud-rate between the main board and the sensor board and reducing the length of the cable to the minimum acceptable size.\n\u2022\tI2C communication problems between the sensor microcontroller and the main microcontroller of SNIFFER device were solved. \nFinally, a USB communication protocol was developed to implement a tablet user interface developed by CEA.\n\nWP5 \u201cBiosensor Development\u201d focused on the development of the sensor sub-module, involving the development of single microcantilever diamond sensors combining Ligand-Binding proteins  selective coatings. \nDuring the first 18 months period, the work had involved the development of diamond microcantilever transducers with optimum performances in terms of mechanical dynamic properties. Moreover chemical grafting procedures were developed and characterized in order to immobilize covalently the bioreceptors onto the diamond surface of the transducers. The sensors were integrated into a gas cell V1 specifically developed for the project\u2019s needs. A sensor array including 8 sensors was designed. A strategy was developed to embed high S\/N read out gauges onto the diamond sensor chips and to actuate all cantilevers together using a single piezo-actuation cell. The sensors are held in a gas cell specifically developed for the project\u2019s needs. This cell has been interfaced with the acquisition electronic and was shown to be operational under laboratory conditions. The integration of 8 diamond cantilever sensors grafted successfully with different proteins and integrated into the gas sensor cell constitutes the V1 biosensor.\nDuring the second term of the project, the diamond micro-cantilevers were developed further. In particular, the piezo resistive gauges integrated in the first version of micro-cantilevers features gauge factors that were too low with respect to the target S\/N ratios expected to reach the necessary sensitivity of the sensors. Therefore a new cantilever structure was designed in collaboration between ESIEE-Paris and CEA so as to integrate polysilicon gauges into the diamond structures. The grafting procedures developed for V1 in order to immobilise the proteins over the diamond transducers surfaces were used again as they were found to be very successful in V1. But this time new proteins with better affinities with target narcotics and explosives were grafted on V2 cantilevers, following the development work on proteins in WP6, in particular by UNIPD, UNIMAN and GTP. In parallel to the work on sensors, a new gas sensor cell was designed based on the original development of the V1 cell. The performances of the V1 cell were acceptable but the major drawback was the fabrication cost associated with the complexity of the design and the resulting machining constrains. Hence efforts were spent on the V2 to simplify the design, without losing performances, in order to cut off the fabrication cost. This was achieved successfully and the V2 version was produced with a reduction of fabrication cost by a factor 4 (based on the fabrication of three units).  The new cell can accommodate 8 sensors in parallel as in version 1.\n\nIn WP6 \u201cProtein Engineering\u201d a first batch of OBPs was produced by UNIMAN. It was composed of four different OBPs (1PDOM from the wasp and F88W from the pig were characterised). Moreover, functional recombinant Major Urinary Proteins MUPs (rMUP 8 and rMUP 20) were expressed by UNIPD. These were implemented in the V1 prototype. Site directed mutations of MUPs have been designed using an in silico strategy. Four target genes have been synthesised. Scale-up procedures and large scale production of 3 OBPs and 2 MUPs were achieved in the first period of the project.  For development of the V2 instrument, six wild type (WT) mosquito (Anopheles gambiae) OBPs, WT Locusta migratoria OBP (LmigOBP1), WTs GOBP2 and POBP1 from Bombyx mori, were expressed and their binding affinities towards the target analytes determined. Three V1 OBPs produced by GTP, WT pigOBP1, mutant PigF88W, and insect wasp 1PDOM, were included. The above OBPs were screened for their binding properties towards V2 target explosives plus some other potential explosives and their derivatives, and they were also screened towards V2 drug targets. In addition, 5 new MUP proteins were synthesised and characterised against explosives and drug. Scale up of production of OBPs and MUPs was carried out by partner GTP. This was not without problems, but for several OBPs quantities between 100 mg to gram quantities were successfully achieved. Selected proteins were then incorporated into the V2 device to produce two types of sensor arrays \u2013 one targeted to explosives and the other to drugs.\n\nWP7 \u201cSelf-diagnostic capabilities\u201d was mainly concerned with the electrical interface and the acquisition of the resonance frequency of each sensor of the array. It also checked that the sensor array was working in the best possible conditions by mean of self-diagnostic and calibration processes. During this first part of the project an electronic architecture had been designed that answers the three main requirements of the work package: readout, calibration and self-diagnostic. \nDuring the first phase an important design effort was made to design an analogue front-end that is both highly sensitive and low noise. The electronic prototype that had been fabricated was able to extract a signal in the \u00b5V range from the 4 micro-cantilever sensors and extract the resonance frequency of each cantilever from it. Results showed that resonance frequency (typically within the range 50kHz \u2013 80kHz) can be obtained with a minimum 10Hz standard deviation under stable conditions. This electronic prototype was fully integrated with the gas analysis chamber developed in parallel in the project. All elements of the signal processing chain had been chosen to that they would  evolve in the second part of the project toward auto-calibration and self-diagnostic. At the same time, equipment had been developed using virtual instruments to serve as a reference characterization bench for the sensors that would be produced during the project. \nDuring the second period of the project, this electronic design had been scaled up to an array of 8 cantilever sensors and design correction and optimization had been carried. The V2 subsystem was answering all the requirement linked to the real-time exploitation of diamond resonant sensors resonance frequency. Important design efforts were made to design an analogue front-end that is both highly sensitive and low noise. The main challenge of this sub-system was tracking the resonance frequency of 8 cantilevers at a time while providing real-time diagnostic based on self-characterization data. The final system was able to compensate before each measurement for the drift in electrical and mechanical characteristic and provide the user alerts and information when the sensor is operating outside normal conditions. This was thanks to the evolution of the V1 prototype toward a V2 prototype based on digitally controlled analog-components; ie a mixed signal architecture. A a real-time fitting algorithm was integrated into the digital signal processing in order to maximize the immunity of the resonance frequency detection to noise. Results showed that the resonance frequency (typically within the range 20kHz to 150kHz can be continuously tracked with a minimum 1Hz standard deviation. This was in track with the objective set at the beginning of the project. This electronic prototype was fully integrated in the global SNIFFER system. Extensive communication had been done with the partners responsible for integration so that the mechanical and electrical characteristics of the subsystem were compatible with the fluidics constraints. The mechanical set-up of the electronic systems facilitated the access and replacement of the sensor cartridge.  Eventually, innocuous chemical samples suitable for calibration of the sensor arrays were identified and recommended for in-field use. \n\nWP8 \u201cDissemination \u0026 Exploitation\u201d, during the first period of the project, was awaiting the first technical results and trials due at the end of the second project year. Thus it focused on the dialogue with users through the SNIFFER User Advisory Group to make sure that the use cases defined by WP1 and the specifications developed in the technical workpackages corresponded well with user needs. A first meeting was organised to present the project and obtain feedback on the documents produced by WP1. Special dissemination activities created awareness through the set-up of a dedicated website, the preparation of dedicated presentation material and the communication about the project to the scientific community.\nThe second period of the SNIFFER project was the opportunity to fully deploy activities. The project dissemination related activities focused on: \n\u2022\tDeepening the dialogue with users through the SNIFFER User Advisory Group. Continuous feedback and advice were sought through meetings and events.\n\u2022\tFurther raising awareness through the updates of the dedicated public website, participation and talks in targeted conferences, seminars, symposiums and public events (supported by the dedicated presentation and communication material). The reached audience was not only the scientific community but it also included end-users and international \/ European institutions interested in border security issues.\n\u2022\tStrengthening contacts with FRONTEX and clustering with other FP7-SECURITY related projects (e.g. DOGGIES, SNOOPY, SNIFFLES, HANDHOLD) through the representation of SNIFFER at their event (e.g. European Day for Border Guard in Varsaw in May 2014) and their invitation to SNIFFER events.\n\u2022\tThe organisation of 2 public events, at project year 2 (May 2014) and project end (May 2015), focusing on the presentation of SNIFFER results to a broad range of stakeholders. The first public event had a strong dog handler colour whereas the second public event focused on the variety of potential end-users if the SNIFFER devices.\nDuring the second period of the project, exploitation activities were also fully deployed. The exploitation plan was fully established and further updated and improved up to project end while better visibility on project results was made possible with time. SNIFFER related patent, publications, potential commercial products and other dissemination activities are listed in the corresponding document.\n\nIn WP9 \u201cProject Management\u201d the SNIFFER Project Office, provided by ARTTIC, first supported the set-up of the different project management tools and processes whilst CEA-LIST was focusing on the strategic and contractual lead of the project. The internal project website was launched at project start and was extensively used by the partners. Dedicated deliverable review processes involve the project partners in ensuring the technical excellence and the usability of the SNIFFER project deliverables.\nDuring the second period of the project, the SNIFFER Project Management Office continued to support the set-up of the different project management tools and processes. Project management tools established at the beginning of the project were kept available to all project partners on the project private website throughout the second half of the project. A risk register was regularly updated throughout the project. These daily actions were carried out until the end of the project and proved successful in ensuring effective day-to-day operational project management. The SNIFFER consortium\u2019s General Assembly met face-to-face every 6 months as scheduled, and the Executive Board met remotely every month. Specific project management issues were discussed, especially that leading to Grant Agreement amendments such as the refining of project scope and project extension, as well as changes in the consortium.\n2.2\tProject objectives for the period \nThe objectives and the corresponding results for the second work period were:\nWP1 Security Usage cases, metrics and validation, societal and ethical aspects\n\u2022\tProvide extensive definitions, user requirements and desired specifications for the final proposed artificial sniffer devices (DL111 \u0026 DL121)\n\u2022\tDefine the methodology for measuring and validating the results from the operational point of view (DL111 \u0026 DL121)\n\u2022\tIntegrate findings and requirements with WP2 and plan the application demonstrator (DL122)\n\u2022\tAddress Societal and Ethical Aspects of such devices and its use (DL111, DL121 \u0026 DL131)\nThey were all achieved as shown in the corresponding deliverables, especially those mentioned into brackets. \nWP2 SNIFFER Device Integration\n\u2022\tV1 system design completed. All subsystem components (sampling, preconcentration and detection) designed, assembled, tested and ready for integration. V1 Integration work initiated.\n\u2022\tV2 submodules  completed and integrated \nResults for both objectives are specifically shown in DL 242 and DL 262.\nWP3 Sub-system for sampling, pre-concentration and pre-treatment of target analytes\n\u2022\tMEMS-based vapour preconcentrators available for integration into V1\/2 SNIFFER systems.\n\u2022\tElectrostatic particle precipitators (ESP) modules available for integration into V1\/2 SNIFFER systems.\nWP4 Sensor Sub-System \u0026 Array Engineering\n\u2022\tDefine of the functional specifications of the sensor sub-module prototype (v1 and v2)\n\u2022\tDefine of the architecture of  the sensor sub-module prototype (v1 and v2): Major changes carried out with respect to v1 to address problems encountered for v2 to produce a robust v2 system.\n\u2022\tLow level electronic and firmware interface and selection of microcontroller for low-level operations: The same microcontroller technology adopted for v1 was used. Minor variations in the firmware were introduced to control the sample inlet temperature and to implement a touch screen display (mounted on the sensor device) and a tablet user interface.\n\u2022\tSelection of pneumatics between sensor array and the other sub-modules\n\u2022\tA neural network strategy based on a radial basis function architecture for programming into a microcontroller environment: No variation with respect to V1\nAll these objectives were achieved.\nWP5 Biosensor Development\n\u2022\tImmobilise OBPs and MUPs onto diamond surfaces + transfer immobilisation protocols onto sensor active surfaces\n\u2022\tOptimise diamond microcantilever transducers in terms and sensor actuation and read out\n\u2022\tValidate the whole sensor principle\n\u2022\tV2 of microcantilever including new read-out gauges and structure (DL522)\n\u2022\tV2 of biosensors including the new proteins developed throughout the project and design and fabrication of new gas measurement cell (DL523)\nWP6 Protein Engineering\n\u2022\tSelection of proteins. Tailoring of protein recognition site conformation for selected analytes through selective mutagenesis (V1 and V2 proteins): Achieved, New OBP and MUP proteins were developed during the second part of the project targeting specific analytes.\n\u2022\tProduction of proteins by bacterial expression systems  and the industrial scale-up of  production of proteins: Achieved for both OBPs and MUPs, some proteins were problematic to scale up production.\n\u2022\tImmobilisation of proteins on to diamond coated cantilever substrates: Achieved, robust coverage of proteins on diamond surfaces was obtained.\n\u2022\tTesting of sensitivity and selectivity of immobilised proteins to selected analytes: Achieved, a range of analytes were tested.\n\u2022\tValidation of stability, repeatability and influence of environmental factors: Achieved, data indicate good stability over time.\nWP7 Self-Diagnostic Capabilities\n\u2022\tDefine a high sensitivity and agile electronic mixed-signal front end architecture specifically for the diamond microcantilever transducers\n\u2022\tDesign, build and verify an electronic board prototype able to extract the resonance frequencies of multiple transducers around the gas cell developed within WP5\n\u2022\tDevelop a reference instrument based on virtual instrumentation to serve as a reference characterization platform for the sensor manufactured within the project\n\u2022\tImplement innovative calibration and characterization process to provide reliable and informative operation of the array of diamond resonant sensors.\n\u2022\tDevelop and implement digital signal processing tools for pushing frequency detection of resonant diamond cantilever to lab instrumentation level within a mobile, autonomous and field efficient device.\n\u2022\tDevelop and implement digital signal processing tools for pushing frequency detection of resonant diamond cantilever to lab instrumentation level within a mobile, autonomous and field efficient device.\n\u2022\tSensor Calibration Methods: Chemicals selected for in-field sensor calibration\nAll the above listed objectives were achieved.\nWP8 Dissemination \u0026 Exploitation\n\u2022\tPromote SNIFFER project and European RTD efforts and make research results available to a larger community: Achieved,  through the set-up of the communication tools \u2013 public website, leaflet, poster \u2013 and presentation of the SNIFFER Project at various external conferences\n\u2022\tFollow up decisions from advisors: Achieved, Regular and valuable UAG and various end-users input all project long.\n\u2022\tPrepare exploitation of SNIFFER results: Achieved through the establishment of an exploitation plan, updated along with project results\n\u2022\tDevelop links with third parties: Achieved, High level of interest from third parties and take-up of the results in a H2020 follow-up project\nWP9 Project Management\n\u2022\tEnsuring the strategic, financial and contractual management of the consortium: Achieved\nAll contractual documents were established, shared and implemented all project long. Needed amendments process was successfully covered by the project management team. Strategic issues and changes in the consortium were dealt with for efficient project management and resulted in a successfully managed project.\nManagement procedures and quality plan were defined and agreed upon at project start. They were implemented all project long. Project management tools were easily accessible and available for all in their latest updates on the project private website. A risk register has been established and regularly updated according to the life of the project. \n\u2022\tSetting up the management infrastructure: At project start, the SNIFFER consortium set up its General Assembly, which met twice a year for a consortium meeting, and its Executive Board, which met every month for a project management follow up teleconference. Each project management issue was dealt with at the appropriate project level, with respect to the project hierarchy.\nProject Results:\n3.1\tWP1 Security Usage Cases, Metrics \u0026 Validation, Societal \u0026 Ethical Aspects\n\nWP number\t1\tStart date or starting event:\tM1\nWP title\tSecurity Usage Cases, Metrics \u0026 Validation, Societal \u0026 Ethical Aspects\nActivity type\tRTD\nParticipant ID\t3\t7\t12\t13\t2\nPart. short name\tST(SI)\u00b2\tCSSC\t3DSA\tINP\tUNIMAN\nPM total (for 60M)\t0\t14\t30\t7\t0.80\nPM spent in period 1\t0\t5,50\t15,10\t4,32\tN\/A\nPM spent in period 2\t4\t0\t12,90\tmissing\t0.40\nPM spent total\t4\t5,50\t28\tmissing\t0.40\nOverview of the progress of the work during the period\nThe objective of WP1 was to provide the necessary Usage Case (UCs) definitions in 3 distinct UCs (according to the amendment to the original DoW). \nIndeed, four devices were initially envisioned to cover the needs of these UCs although the core modules in all devices will be the same:\n1.\tSNIFFER BODY SCANNER: A portable scanner able to identify illegal and\/or dangerous substances (like drugs and explosives) by detecting their explicit odour. The first envisaged device is a portable (ideally handheld) body scanner that will be used to scan people. The procedure is almost identical to the common procedure of testing a person with a metal detector. That is why ideally this scanner should be integrated into a metal body-scanner, standard equipment in many security sensitive environments (critical infrastructures, airports, etc.).\n2.\tSNIFFER PORTABLE SCANNER: A device that will be able to identify illegal and\/or dangerous substances (like drugs and explosives) by detecting their explicit odour. The envisaged device is a portable (ideally handheld) scanner that will be used to scan open luggage and cars (interior).\n3.\tSNIFFER PORTABLE CONTAINER SCANNER:  A SNIFFER device will be able to detect clandestine and illegal immigrants, hiding in a container or a truck by detecting their explicit odour. The envisaged device is a medium size scanner that will be used to scan containers.\n4.\tSNIFFER PORTAL SCANNER:  A SNIFFER device will be able to identify illegal and\/or dangerous substances (like drugs and explosives) by detecting their explicit odour. The envisaged device is a portal size scanner that will be used to scan luggage on a running belt.\nFollowing the project mid-term review, the original list of Usage Cases has been reduced to three concrete usage cases:\n\u2022 UC3: controlling suspicious open luggage and cars with small portable scanning device (V1)\n\u2022 UC4: controlling suspicious open luggage and cars with small portable scanning device (V2)\n\u2022 UC6: systematically scanning closed luggage on a running luggage belt for illegal substances and explosives (V2). The above corresponding devices covered all described UCs and served as a basis for WP2 and the rest of the technical development.\n\nWork progress in each task \nTask 1.1 \u0026 1.2 \u2013 UCs for V1 and V2 technologies\nCSSC\u2019s initial contribution consisted in a preliminary analysis of the ethical and societal implications generated by the perspective deployment of SNIFFER devices for security checks at borders and at other critical installations. It also consisted in the elaboration of recommendations for field trials. The above preliminary analysis focused on 4 devices for 6 UCs: UC1 \u0026 UC2, UC3 \u0026 UC4, UC5, UC6. Unfortunately, CSSC went through voluntary liquidation early 2014. However, the burden of ethics was lessened following the first review meeting as the scope of the project was re-focused. \nMoreover, the French ministry of interior as an end-user participated in the elaboration of the UCs. The ST(SI)\u00b2 with the support of CREL from the national police made a field inquiry in the airports with the different bodies in charge of security: civil aviation, specialized gendarmerie, customs, border police, ... The CREL made a great job in the technical definition of the devices that could be developed. It managed to get some specifications from various end-users who shared main practical constraints in the field of detection of dangerous substances on the borders. The CREL prepared a methodology for the future laboratory tests and submitted it to the consortium. \nINP\u2019s contribution consisted of forming a workgroup with relevant Police Departments such as: The Operations Department, The Department of Investigations \u0026 Intelligence, and The Police Legal Adviser. It collected information and requirements from end-user\u0027s in the Police Department regarding various aspects such as legal requirements, operational requirements, ethical requirements. It also contributed to the planning and implementation of trials (field trials and lab trials of TTS machine): testing requirements, sites for testing, and legal requirements.\nAfter the 1st Project Review and following recommendations and discussions with the PO and reviewers, it was decided to limit 3 Use Cases as stated above. UC3 lead to a lab trial, which allowed the SNIFFER users to assess under lab conditions the dedicated devices delivered by WP2. UC4 and UC6 in turn lead to field trials, which allowed the SNIFFER users to assess under close to real life conditions the dedicated devices delivered by WP2. Each Usage Case task followed the same \u0022definition - trial - assessment\u0022 cycle, lasting 24 months and starting at T0 (T0 =Month 1 for the first usage case and T0 = 13 for the second usage case):\n1. Define the usage case from the different perspectives (T0 - T0 + 4 months)\n2. Regularly stay in contact with users to present progress, obtain additional inputs (information, samples, etc.) and plan deployment of trial (T0 + 9 months; T0 + 26 months; T0 + 37 months)\n3. Set up the trial: Receive devices, install, train users, setup demonstration (T0 + months 18 to 25) (T0 +months 25 to 37)\n4. Run the field trial, assess and write assessment report (T0 + month 40) with support from technical partners in WP2.\nThe target compounds for the lab tests and field trials will be limited to the following:\n\u2022 Explosives: 2,4 DNT, 2,6 DNT, and TNT are required. RDX and PETN are the priorities. HMX, EGDN and ammonium nitrate are optional for V1 and will be included for V2.\n\u2022 Drugs: Cocaine, Ephedrine, Heroin, THC, Ecstasy (MDMA)\n\u2022 Other: Atropine (to show the ability to detect other substances)\nThese tasks were completed with the submission and acceptance of the revised deliverables that limited the use cases to 3 and described a more viable approach to the project.\n\nTask 1.3 \u2013 Collection of ethical approvals\nCSSC initially assisted WP1 leader in the drafting and reviewing of a detailed information package to be sent to National Data Protection Authorities that describes the project, its main mid-term results and presents the procedure to be followed to organise and implement field trials. In particular, CSSC contributed to the document by presenting an overview of the EU privacy and data protection legal framework and reaffirming the ethical standards to be followed during the field trials that will be held in Greece and Israel. Unfortunately, CSSC went through voluntary liquidation early 2014. However, the burden of ethics was lessened following the first review meeting as the scope of the project was re-focused. \nDeliverable 131 \u201cReport on National Ethical Approvals\u201d was rewritten and resubmitted on Dec 2013. That deliverable covered the EU privacy and Data Protection Legal Framework and presented the outline for the Trials Procedure Protocols. 3DSA was also appointed (as the WP leader) to investigate the required actions and necessary authorisations for the implementation of the final field trials according to the trial protocols that would be developed. \nTo simplify things and avoid time consuming authorisations, the consortium decided to restrict the field trials to the members of the consortium and a few stakeholders (Greek Police, Athens International Airport authorities). This decision eliminated the need for authorisations. Nevertheless 3DSA contacted the Hellenic Data Protection Agency (HDPA) and after some communications and presenting of the case, HDPA ruled that authorisation was not necessary and field trials could take place as planed.\n\nTask 1.4 - Independent Ethics Advisor\nProf. Timothy Stibbs from the University of Manchester, an experienced ethics specialist, oversaw with impartiality the ethical concerns raised by this research involving human beings.\nPlanning of experiments involving public venues and reports associated were screened by the ethical advisor. Mid-term by the Ethics Advisor were submitted to the EC\/REA with the Periodic Report. The same process will be followed concerning final reports.\n\nSignificant results: \n\u2022\tPreliminary ethical and fundamental rights analysis of the UCs together with recommendations for field trials\n\u2022\tSuggestions for modifications and adjustments to suit the real operational requirements and legal limitations\n\u2022\tEstablishing lab trials prior to field tests\n\u2022\tSet up and organisation of the Final Field Trials at Athens International Airport\n\u2022\tSuccessful delivery of all WP DLs, specifically resubmission of DL111 and DL 121, as well as DL 131 according to recommendations from the 1st review report (Dec. 2013)\nDescription of problem \/ deviation\tSuggested \/ undertaken corrective action\tImpact on other WPs, budget\/PM plan or work plan\nLack of information concerning the prototypes to be tested at V1 field test scheduled for UC1 and UC3, lack of reports (technical + operational + safety)\tRequirements have been sent \u2013 ongoing reply\nLab\/field trials technical meetings (June and September 2013)\tDelayed planning of field trials.\nNot able to find narcotics at the Athens International Airport Trials (Airports are not interested in these)\tTrials took place with explosives. Samples were more than enough to test the protocols and equipment.\tNone as extensive lab trials have been ran with the above list of narcotics. \n\n3.2\tWP 2 SNIFFER Device Integration\nWP number\t2\tStart date or starting event:\tM1\nWP title\tSNIFFER Device Integration\nActivity type\tRTD\nParticipant ID\t2\t5\t1\t11\t\nPart. short name\tUNIMAN\tAIRBUS\tCEA\tTTS\t\nPM total (for 60M)\t30\t15,50\t17,40\t3\t\nPM spent in period 1\t9\t10\tN\/A\tN\/A\t\nPM spent in period 2\t22\t2,16\t17,40\tmissing\t\nPM spent total\t31\t12,16\t17,40\tmissing\t\nOverview of the progress of the work during the period\n\u2022\tWP2 achievements to M4 was related to detailed specification and definition of functionalities and system design architecture for SNIFFER devices 1 and 2 focusing on body scanning devices and searching technology at open luggage and cars.\n\u2022\tUp to M18 a more detailed specification and description is supplied in device 2, V2 developments. Deliverables 5-6 were focusing on UCs 4-6, which deal with container inspection and luggage belt screening in airports.\nWork progress in each task \nSystem design in sampling, preconcentration and detection device developments (V1+V2) are finished at the end of reporting period. All related and relevant mechanical and electronic sub-elements are constructed, integrated and tested already.\n\nTask 2.1 \u2013 Develop D1 \u2013 SNIFFER V1 of an enhanced metal body-scanner with \u201cartificial sniffer\u201d functions\nThis task was cancelled with the amendment of the SNIFFER Description of Work at mid-project.\n\nTask 2.2 \u2013 Develop D2 - SNIFFER V2 of enhanced metal body scanner with \u2018artificial sniffer\u2019 functions\nThis task was cancelled with the amendment of the SNIFFER Description of Work at mid-project.\n\nTask 2.3 \u2013 Develop D3 - SNIFFER V1 of a light portable device for open luggage and cars\nD3 device architecture was defined in the same way as that developed for D1. The main difference between D1 and D3 is in the sampling flows. The consortium selected stainless steel tubing and valves for this prototype. The electronic board for D3 was developed, built and tested. It is the same as for D1. In fact, it allowed selecting the right flows required by this different application. USB, SPI, I2C, Ethernet communications system were integrated the system for communicating with the user and the different submodules.\nKeypad and LCD were also implemented as an easy user interface.\n\nTask 2.4 \u2013 Develop D4 - SNIFFER V2 of a light portable device for open luggage and cars\nD4 detector sub-module was modified respect to the V1 version to improve sensor performance. \nThe device was conceived as an independent sub-module that can be controlled and run by an end-user. When sample detection needs to be performed, the user selects start measurement on the touch screen display and the following procedure is started:\n1.\tThe Main microcontroller sends a signal to the thermo-desorber\/pre-concentrator communicating that it is ready for starting a measurement;\n2.\tThe thermo-desorber\/pre-concentrator notifies the detector that it is ready for desorbing the sample and then wait for the start signal from the detector;\n3.\tThe detector sucks air at 100 sccm from the inlet for 10s to record the baseline and then sends the \u201cstart desorption\u201d signal to the desorber\/ pre-concentrator;\n4.\tThe sampler\/ pre-concentrator starts the desorption while the detector is sucking at 100 scmm. Hence the vaporised sample is delivered to the biosensor. The proteins bind the analyte molecules changing their shape. The transducer converts the binding effects into a variation of the sensor electronic signal read. This step last for 40 sec;\n5.\tThe thermo-desorber\/pre-concentrator communicates to the master the end of the desorption;\n6.\tThe detector sucks air from the inlet to clean the sensor;\n7.\tThe sensor data are immediately processed by the radial basis neuronal network software and the results are shown on the touch screen display or a HMI.\nThe first v2 ESP sampling development (DL321-324) worked effectively and fulfilled all needed requirements. Following several trials it was decided that for Athens Airport field tests a new more practicable device configuration should take place. A handheld sampling device D4 based on corona discharge and electrostatic precipitation of particles targeting sampling of narcotics or explosives with extremely low vapour pressures was developed and delivered on time. The physical principles behind this technology and the technical details of this device can be found in DL242. The device was assessed in laboratory conditions and then in Athens International Airport to sample illicit particles e.g. in passenger open luggage. This sampling device D4 is also suitable for searching cars in particular for particles of narcotics and eventually explosives.   \nThe new configuration in D4 and D6 device operations also required also new thermos-desorption figures which are described more in detail under DL321-324. A new thermos desorption device was designed and configured with regard to the needed operational procedures and figures. \n\nTask 2.5 \u2013 Develop D5 - SNIFFER portable device for containers\nThis task was cancelled with the amendment of the SNIFFER Description of Work at mid-project.\n\nNB: A macro preconcentrator that can sample large volumes of air was developed by partners ARMINES and EPFL and interfaced to the SNIFFER detector submodule. Such device is particularly suited for search of volatile odorants such as tobacco odours in containers. For obvious practical reasons the device was tested in laboratory conditions for the detection and identification of tobacco.  \nTask 2.6 \u2013 Develop D6 - SNIFFER Monitoring Portal for a luggage conveyor belt\nWhile D4 needs to scan open luggage, D6 aims at controlling close luggage on a conveyor belt. Hence, they have been designed as two devices having the same sensor array but with a different sampler.\nAn automatic Trace Collector module was developed by partner TTS for searching particles of narcotics and explosives in closed luggage. The module was tested and validated in Athens International Airport in combination with the other SNIFFER submodules successfully. Although the device was not directly tested in conveyor belt configuration the technology could easily be adapted to such application via reasonably low further engineering efforts. Similar technology has already been demonstrated by partner TTS in conveyor belt configuration. Indeed, it was decided that D6, being only a technology demonstrator did not require the use of a conveyor belt.  As a consequence, the device was to be built without one. The requirements for the machine were to be able to accept the largest of hand-carried baggage typical of air travel and extract traces of materials on and within the bag for purposes of Electronic Trace Detection (ETD) methods to be applied to the traces. TTS took the lead for the design and production of this device as it is the world recognized leader of this technology. The design included application of TTS\u0027s technology for using aerodynamic flows to accomplish the extraction and collection of the traces.  A chamber was devised, somewhat larger than the largest hand-carried baggage in which the aerodynamic devices were placed. The frame and supporting controls and air handling equipment was procured and the device assembled.  The device was tested using a number of powders as simulants as well as laptops and few bags to determine the collection capability.  The device passed these tests even without optimization.  After a few months, the device was taken to the Israeli National Police for testing with live explosives and narcotics.  It was fortunate that the test facility was near the INP canine facility and side by side testing was done.  Troubles with the air providing devices prevented a full set of tests, but the those accomplished showed that the D6 device was capable of extracting materials even from within closed bags.  In this series, the canines proved extremely reliable as well.\nSubsequent testing at the Athens International Airport in Greece again proved that the D6 device was capable of extracting even small amounts of explosive materials from within bags which had been zipped close.  The amount of materials extracted were sufficient in every case for the SNIFFER device to detect their presence.\n\nTask 2.7 \u2013 Final Consolidated Technology Report\nBetween the last sampling V2 developments and at the beginning of the field trials at Athens Airport some issues with regard to maturity levels and operational behaviour were identified. Most of them could be solved and at some degree the outcome solution showing up better results as it were expected at the beginning. \n\nSeveral outcomes and problems encountered between our last V2 sampling-desorption elements could be discussed in following manner:\n\nProblems regarded to operational procedures\n\u27a2\tOriginally ESP sampler wasn\u2019t small and portable enough and needs power line interconnection, which shows up not very convenient and valuable for the field operation.\n\u27a2\tNon testability of TTS luggage belt sampling system bevor the field trials\n\u27a2\tDifferent sampling pad sizes and materials from luggage belt and thermos desorption and ESP sampling devices\n\nProblems encountered after laboratory trials\n\u27a2\tStill contamination of glass inlet tube from desorption chamber\n\u27a2\tSolenoid thermo-desorption system fails because of long term overheating and connectivity problems\n\u27a2\tHeat stamp cables insulation were scratched after several ours operation time \n\u27a2\tInterface needed from V1 sampling system wouldn\u2019t be appropriate for the new design\n\nProblems with regard to the timeline\n\u27a2\tVery short timeline between laboratory trials and field test\n\u27a2\tNew configuration of interface and controller occupies great amount of time\n\nSolutions:\n\u27a2\tHeating of inlet block and transfer-lines of sampling inlet\n\u27a2\tDesign and build new ESP handheld sampling device\n\u27a2\tThermo-desorption-heat protection screen with perforated plate for safe operation\n\u27a2\tRefurbishing of heat stamp because the coating of cables were damaged\n\u27a2\tCreative and intense communication between WP3 and TTS solving probably interface issues solved very successfully\n\u27a2\tCreation of new desorption communication and control interface\n\u27a2\tTTS designed new filter holder and also tested our envisaged filter pad materials to fit into the needs with regard to the different devices and usage cases.\nSignificant results: \nMajor system integration aspects between partners were gathered. A systematic device integration concept of functionalities and integration aspects could be delivered. Functionalities are aligned and systematic interaction in device integration is at a mature level for device integration 1 to 4.\nAll SNIFFER submodules V2 were delivered and finally integrated on time to be tested. Following laboratory validation, the SNIFFER detection module, the Automatic Trace Collector sampling module and the Electrostatic particle collector were tested together in the same environment in complementarity for the first time in Athens International Airport, in an operational environment: the airport luggage-check area. The SNIFFER results were benchmarked against an IMS (state-of-the-art in airports today). The sampling systems proved to be efficient in collecting particles hidden in back bags (validated by IMS first then SNIFFER): 100 % of hidden particles were detected and therate of identification was assessed to be around 60 %.  Those results are very promising and were well received by end-users and stakeholders to whom they were presented at the subsequent SNIFFER workshop.\nDeviations: None\n\n\n3.3\tWP 3 Sub-system for sampling, pre-concentration and pre-treatment of target analytes \nWP number\t3\tStart date or starting event:\tM1\nWP title\tSub-system for sampling, pre-concentration and pre-treatment of target analytes\nActivity type\tRTD\nParticipant ID\t4\t5\t6\t11\t\nPart. short name\tARMINES\tAIRBUS\tEPFL\tTTS\t\nPM total (for 60M)\t26\t17\t36\t16\t\nPM spent in period 1\t15\t10\t15\t8,75\t\nPM spent in period 2\t9,03\t10\t18\tmissing\t\nPM spent total\t24,03\t20\t33\tmissing\t\nOverview of the progress of the work during the period\n\u2022\tA particle sampling system using electrostatic precipitation (ESP) technique was developed. Sampling tests showed that the system is able to sample selectively allowing preconcentration of target substances. \n\u2022\tThe ESP approach will be further improved by modifying the sampling electrode (ceramic chip) by adding a conducting layer on top which allows electronic discharge when particles are sampled. This was done using aluminum deposition techniques. Tests will be carried out in the next weeks.\nWork progress in each task \nAn electrostatic particle precipitator (ESP) sampler unit was initially developed at EADS (former AIRBUS) and integrated into a breadboard system. This breadboard setup allows the ESP sampler to be combined with various kinds of detectors to perform full-chain sampling-detection experiments. The sampler on its breadboard unit allows solid particle residue (drugs, explosives) to be extracted from suspect surfaces through hydrodynamic forces (vacuum cleaner). The sampling air stream is then guided through the ESP sampler where the sampled residue becomes electrically charged upon transit through a corona discharge. Depending on the polarity of the discharge needle, either highly electron (explosives) or highly proton-affine (drugs) substances become electrically charged and immobilized on a heatable ceramic chip. After the sampling step the chip with the collected substance is moved outside the particle sampler underneath the SNIFFER detector entrance port. This method of sample transfer minimizes the loss of target substance due to vapor condensation on  internal walls of the SNIFFER device. At EADS successful full-chain sampling and detection tests have been performed with TNT (explosive) and Ephedrine (drug) particles and a number of common interfering substances (mineral dust, pollens, food stuff, medical drugs). As SNIFFER OBP-detectors had not yet been available, commercial metal oxide (MOX) or surface ionization (SI) detectors had been used in these preliminary experiments. \n\nA news V2 electrostatic particle precipitator (ESP) sampler unit was developed at Airbus with new handheld metrics and figures.  This new setup allows the ESP sampler to be combined with various kinds of detectors to perform full-chain sampling-detection experiments (IMS \/ OBP). The sampler unit allows solid particle residue (drugs, explosives) to be extracted from suspect surfaces through hydrodynamic forces (vacuum cleaner). The sampling air stream is then guided through the ESP sampler where the sampled residue becomes electrically charged upon transit through a corona discharge. Depending on the polarity of the discharge needle, either highly electron (explosives) or highly proton-affine (drugs) substances become electrically charged and immobilized on a heatable ceramic chip. After the sampling step the sampling pad with the collected substance is moved outside the particle sampler and feed into the new OBP-thermos-desorption device. This method of sample allows the transfer either from the ESP sampling system as well as from the luggage belt UC6 set up. At Athens Airport successful full-chain sampling and detection tests have been performed with different explosive particles and a number of common interfering substances. As SNIFFER OBP-detectors had not yet been tested under real environmental conditions all test was compared as well with IMS detection devices as well. \n\nHandheld aspiration device for the final field tests at Athens airport V2\nA more flexible and more \u201chand held\u201d based sampling system for the field test with regard to the use case at the Athens Airport was developed (V2) with mayor focus of operability with regard to the use cases. \nSustaining this needs and reducing cost and time we decide to use a commercial of the shelf (COTS) -based vacuum cleaner system in which the required ESP mechanical and electronically parts could be integrated.\n \nOverview of Handheld Sampler assembly, ESP with particle aspiration and air-sampling system\n\nESP-Handheld Device (V2) - Structural Changes and New Features\n\u27a2\tFollowing field testing at AIA, the ESP-sampling module needs optimization to be suitable for the operational procedures needed for luggage and car inspections.\n\u27a2\tFollowing lab test, V1 was already rearranged hand-held and working independently from the lab instrumentation (power supplies, measurement instrumentations...)\n\u27a2\tUsing a battery (NiMH 14,4V) driven vacuum blower configuration to subtract the particles form surfaces of investigations (e.g. seats, luggage, cloth..)\n\u27a2\tNew design to be based on COTS-based instrumentation including all SNIFFER new developments for a more comfortable usage in conjunction with or operational requirements\nInertial Impactor practical Sniffer-Handheld-ESP-Sampling-Design\nIn our new handheld based ESP sampling approach the introduction of the impactor technology shows up in a higher sampling efficiency and leaving out most of the unwanted bulkier particles originated from dirt and other air impurities.\nTo our knowledge this is the first time that the impactor and ESP technology were integrated into a particle sampling system for security screening approaches.\nThe figure below sketches the sampling- air pathway until it reaches the ESP separation chamber. It could be seen that the larger sized particles (purple) are passed away before charging, only the smaller sized one can reach the area of ionization needles.\n\n \nPre-separation impactor-chamber for extraction of bulky and heavier particle fractions from air streams\n\nTask 3.1 \u2013 Development of the testing bench for DNT generation and TPD studies\nARMINES\u2019s contribution mainly focuses on the implementation of a test bench and the optimization of adsorbent materials to fill up the preconcentrator devices whereas EPFL efforts were centred on the microfabrication of silicon and plastic preconcentrators. \nTherefore the test bench implemented in project is basically composed of a gas generator from Owlstone Nanotech using permeation tubes, mass flow-meters and a climatic chamber containing the preconcentrator device and the detection system. The climatic chamber is maintained at 40\u00b0C to avoid any condensation between the gas preconcentrator  and the detector. Different permeation tubes (TNT, DNT, others interfering gases) were purchased and calibrated by weight loss at a given temperature for estimating their permeation rates. According to the vapor pressures, the maximum concentrations which can be obtained at 25\u00b0C are 8 ppb for TNT and 482 ppb for DNT.\nMeanwhile Four adsorbent materials: a carbon nano-powder, TENAX powder, porous silicon and Carbopack B were characterized using TPD (Temperature Programmed Desorption) and BET (Brunauer, Emmett and Teller) method for evaluating their affinity toward target gases in the SNIFFER project (explosives and drugs) and also for estimating their desorption temperature. Globally these experiments show high desorption temperature of explosives (desorption between 350 and 400\u00b0C) on carbon nano-powder and on TENAX while porous silicon and on carbopack B show low desorption temperature (down to 250\u00b0C) of DNT compounds. For this purpose, carbopack B and porous silicon were chosen for the preconcentration of explosives and Tenax is used for the preconcentration of drugs derivatives such as acetophenone.\nThree versions of gas preconcentrators were developed through SNIFFER. At the beginning of the project a micro preconcentrator based on silicon technology was implemented for fulfilling the requirement of small size device for the version one of Sniffer handheld detector. However, our partners suggested the development of macro-preconcentrator for large volume sampling for the user case of cargos inspection. Following the development of the macro-preconcentrator, another small size device based on plastic technology was implemented in order to ally small size and large volume sampling since this device possesses adjustable inlet\/outlet allowing high flow rates.\n\nSilicon micro gas preconcentrators\nTwo different micro-preconcentrators have been developed during the sniffer project. The first ones are based on silicon technology. They are made with a succession of photolithography and DRIE steps and bonded with a glass cover by anodic bonding. The design of channels within the device and the inlet\/outlet vary according to the adsorbent material and the gas to be detected. Some have a preconcentration chamber for receiving a large amount of adsorbent and some are made with a main channel for maximizing the probability for a gas to be detected. Silicon micro gas preconcentrators include a platinum heater at the backside deposited either by lift-off process or by screen printing. Prior the heater deposition, the backside of the micro-devices is coating with 150nm of Si3N4 in to avoid any electrical breakthrough during desorption over 300\u02daC.  \nMeanwhile, the use of porous silicon as an adsorbent or\/and a binding surface was also studied. Indeed, pores were grown within channels and in the preconcentration by anodic etching in HF bath. Besides enhancing the preconcentration performance of the device, the pores allow easy desorption of target gases by lowering their desorption temperatures. \nLong used in SPME (solid Phase Micro-Extraction) fibers for gas chromatography, Carbopack B has been chosen as adsorbent for also its affinity towards target gases in SNIFFER context such as DNT, BTX (Benzene, Toluene, Xylene) and Acetophenone. Nevertheless, others adsorbent like carbon nanopowders and Tenax were considered. Adsorbents were inserted in micro-devices in granular form before the sealing of glass cover for preserving its interstitial porosity and thus its high surface to volume ratio.\nAfter the adsorbent deposition, the micro gas preconcentrator is then decorated with micro-capillaries as fluidic connections. In the first proposed designs the fluidic capillaries are laterally connected to the device whereas the new design of micro gas preconcentrator  offer the possibility of using large capillaries since inlet\/outlet are directly dug in the glass cover (Figure 1). The devices allow flow rates up to 20L\/h with a pressure drop down to 2bars and could reach a heating rate of 60\u02daC\/s thanks to their small thermal mass.\n\na)   b)   c) \nImages of a) First design of silicon preconcentrator at the final stage with lateral fluidic connections and heating element b) Second design of silicon device platinum heater at the backside c) 3D view of the second design with fluidic connections on the glass cover.\n\nIn overall, the combination of the new preconcentrator design with porous silicon and carbopack B shows the highest preconcentration performances with a complete desorption in comparison with the old design and\/or without porous silicon when tested at the same experimental conditions.\n\n    \nDNT peak after an adsorption of 200ppb@5min at 20L\/h and a desorption at 250\u02daC with a flow of 10L\/h obtained with a old design (left) and new design (right) of gas preconcentrators with filled 1mg of carbopack B.\n\n \nComparison between carbopack and Porous silicon tested separately and in combination after adsorption of 200ppb@5min of DNT at 20L\/h and a desorption at 250\u02daC with a flow of 10L\/min.\n\nMacro-preconcentrators\nThe macro-preconcentrator is basically made with a metallic tube filled with carbopack B and wrapped up with a heating element. The macro-preconcentration is incorporated with pump for defining flows, a fan as a cooler and PCBs for obtaining an autonomous operating system. A photograph of the macro-preconcentrator system is shown in Figure 4. The submodule can work either in compression mode or vacuum mode (Figure 5). In the compression mode the pump is upstream the pre-concentrator; this mode allows the higher flowrates up to 18 L\/min, but there is a risk of contamination by the pump (even if the pump is a \u201clow pollution\u201d dry piston pump). In the vacuum mode the pump is downstream the preconcentrator and then sucks the gas in.  This mode eliminates any risk of contamination but allows lower flowrate (8 L\/min). The current system is delivered with the vacuum mode. Tubing modifications could allow switching easily to the compression mode.\n     \nPhotographs of the macro pre-concentrator\n\n           \nTwo modes of operation of the macro-preconcentrator submodule. Right: Compression mode, Left: vacuum mode\nThe macro-preconcentrator device was tested by our CEA partners under toluene and cigarettes. These results have shown an increase of the sensitivity of SAW sensors array when operating with a preconcentration at different adsorption time (figure 6). Besides, the pattern recognition algorithm developed in Sniffer project allows separating aromatic compounds such as toluene from cigarettes which is one of the main interfering compounds in context of explosives and drugs detection (figure 7). \n  \nResponse of an array of SAW sensors toward toluene (left) and (cigarettes (right) adsorbed at 450mL\/min and desorbed at 250\u00b0C with a flow of 350mL\/min.\n  \nFingerprint of toluene (left) and cigarettes (right) with an array of 8 SAW  sensors for adsorption time of 10min.\n\nFoil gas preconcentrators\nThe second small devices developed in this project were based on plastic foil technology.  These foil gas preconcentrators are made by rolling up a gold micro-hotplate printed on foil substrate before filling it with the desired adsorbent. This concept allows easy fabrication of gas preconcentrators with adjustable inlet\/outlet diameter depending on the adsorbent to insert and the flow rate to be imposed. Two chosen adsorbents (carbopack B and Tenax TA) were inserted within the rolled hotplates in granular and blocked with glass wool. The fluidic and electrical connections are then sealed with ceramic cement and conductive glue, respectively (Figure 8). Foil devices are usable at flow rate up to 100L\/h and at temperature up to 350\u02daC.\na)  b)   c)  \nImplementation process of the foil gas preconcentrator: images of a) a printed hotplate with electrical wires b) The rolled up micro-hotplate compared to 5cent and c) The final Foil preconcentrator.\n\nThe foil gas preconcentrators are successfully tested under benzene, acetophenone and nitrobenzene (figure 9). These results show preconcentration performances better than silicon devices thanks to the use of high adsorption flow rates, up to 80L\/h (20L\/h for the silicon). \n\n    \nSensor response for a foil gas preconcentrator filled with (left) 1mg of carbopack B when exposed to250ppb@1min of benzene at 40L\/h and desorbed at 200\u00b0C with a flow of 4L\/h and right) 1mg of Tenax TA when exposed to 100ppb@10min at 20L\/hof acetophenone and desorbed at 130\u00b0C with a flow of 2L\/h.\nIn addition, foil micro gas preconcentrators permit also operating with a duty cycle down 30s (crucial in case of explosives detection) while conserving its high preconcentration performances (figure 10). \n\n \nBenzene peak for a FGP filled with 1mg of carbopack B after a adsorption of 250ppb during 60, 10 and 5s at 40L\/h and desorbed at 200\u00b0C with a flow of 4L\/h.\n\nTask 3.2 - Design and fabrication of new devices (IMT4) and feasibility study on flexible substrates \nConcerning the development of pre-concentrator devices on plastic, preliminary temperature tests are currently realized by TPD in order to determine the range of temperatures which can be used with the different polymers. On the other hand, for the OBP deposition in the device, it is expected to use inkjet printing. The adaptability of the inkjet machine of ARMINES and the determination of the deposition parameters are the objectives of the next period.\nNew preconcentration devices (IMT4) have been designed and fabricated in silicon technology. These devices are smaller than the IMT3 ones and have a serpentine design for maximizing the adsorption rate. In opposition to previous devices, IMT4 devices have a fluidic connection on the glass cover which allows no limitation on the capillary size since inlet\/outlet are not dip in the silicon As IMT3, these new micro-channels include also a platinum micro-heater on the backside made by micro-fabrication process (Lift-off). These IMT4 devices are under optimization using FEM model for thermal and fluidic simulation for maximizing the adsorption capacity and the heating efficiency.\n\nTask 3.3 \u2013 Module for automatic trace collection on hand-luggage for device D\nFurther Development of Trace Extraction Methodologies\nThe current methods of removing particles and vapours from within closed carry bags and from the external surfaces were examined for improvement. For internal extraction, it remains to be sufficient to cyclically pressurize and depressurize a small chamber containing the examined package. Studied was the pressurization amplitude and frequency. A new data base is being developed for various types of packages so that the most effective pressure variation can be part of a program for the operator.  This data base will be used in the future.  For now a very good scheme has been developed for both particles and vapours. For external contamination (particles only), a series of jets have been shown to be effective at loosening pieces of contaminants to join the airborne particles and vapours during the extraction phase. Jet timing and speed are important parameters in this process. \nFurther Development of Trace Concentration Methodologies\nThe particles and vapours extracted from the above phase are now distributed throughout the chamber and must be concentrated and collected.  This should be done thoroughly to avoid chamber contamination in the event of the presence of contamination. Collection and direction methods were improved for particles as small as \u00bd microgram. Specialty jets were developed to take advantage of the special nature of airborne particles as they begin to fall and for the larger particles as they impact the floor and walls of the chamber. The jets cause the particles to be directed toward a target collector which currently is a fiber-glass filter capable of holding particles down to 1 microgram. These particles can later be vaporized if desired to provide vapours from them. Vapours in the chamber naturally follow the exhaust air and are sampled upon exit from the chamber.\n\nDesign of Device for collecting trace vapours and particles from hand-carried \nIn this task a device suitable for the SNIFFER program was completely designed with all engineering work including ergonomic considerations completed. The device includes conveyor means to automate placing the inspected articles in the testing chamber.  For the field trials, it should be easy enough to provide the basic system without the conveyors to save costs.\nTesting of using heating to improve concentration of extracted particles resulted in stickier particles rendering the method ineffective at improving extraction \u003E Limit heating to 35\u00b0C.\nTesting of collection concepts through a diaphragm: the idea resulted in a more complex diaphragm design rendering it more expensive.  It further could not be made to be effective for the all-important larger particles which would get stuck on the walls \u003E Corrective action was to collect vapours and particles from the floor of the chamber. This was the current practice so this was no problem to use.\nSignificant results: \n\u2022\tFlexible preconcentrator + OBPs as an absorbent material\n\u2022\tOptimized devices obtained\n\u2022\tMacro-porous silicon preconcentrator fabricated\n\u2022\tUnder integration in finite device\n\u2022\tOptimization of DNT generation and process for study of adsorbent materials by TPD\n\u2022\tValidation of porous silicon and Carbopack B for the preconcentration of DNT\n\u2022\tValidation of Si3N4 insulating film for heater and process for capillaries sealing\nWith the involvement of a new \u201cview\u201d perspective within integration process a more complete integration progress is coming into progress.\n\u2022\tDevelopment of high reliability method for trace particle extraction developed and for trace vapour extraction;\n\u2022\tImproved aerodynamic means of particle control and concentration and of concentrating vapours;\n\u2022\tComplete system for collecting trace particles and vapours from hand-carried packages designed.\nDeviations: None\n\n3.4\tWP 4 Sensor Sub-System \u0026 Array Engineering \nWP number\n4\tStart date or starting event:\tM1\nWP title\tSensor Sub-System \u0026 Array Engineering\nActivity type\tRTD\nParticipant ID\t2\t\nPart. short name\tUNIMAN\t\nPM total (for 60M)\t21\t\nPM spent in period 1\t9\t\nPM spent in period 2\t1,40\t\nPM spent total\t10,40\t\nOverview of the progress of the work during the period\n\u2022\tThe hardware integration of the biosensors into a small portable sensor sub-module\n\u2022\tThe multi-parametric training of the biosensors to optimise the recognition potential from the sensor array\nWork progress in each task \n\u2022\tAll related DL submitted\n\u2022\tFunctional Specifications and Design of the V2 system was achieved\n\u2022\tParametric software for pattern recognition was written and tested\n\u2022\tV2 sensor-sub-module was integrated and the system went through lab and field trials\n\nTask 4.2 V2 of Sensor Sub-system\nThe work in this work package included:\n\u2022\tThe hardware integration of V2 biosensors\n\u2022\tThe multi-parametric training of the biosensors to optimise the recognition potential from the sensor array\n\u2022\tThe integration of software and hardware into the V2 Sensor Sub-System\nFour deliverables were produced: DL421, which defines the functional specifications and design of architecture of V2 Sensor; DL422, defining the parametric software algorithms for pattern recognition, DL423, aimed at the description of the version 2 of the sensor sub-module and DL424, that reports on technical support and problems encountered during the V2 development.\nThe second version of the Sensor Sub-System was built and integrated. It is composed of 3 main parts: pneumatic system, main electronic board, and sensor system.\nPneumatic system\nV2 pneumatic system was modified respect to the V1 version. The new arrangement is shown below:\n\nThe V2 delivery system is composed of two main parts: Sample pathway and Humidity line. \nHumidity line\nThis line provides on-demand humid air to the biosensors. It is composed of a 2-way valve, a small glass bottle containing distilled water and a nafion humidifier tube and a pump that control the flow through the nafion tube. \nSample Pathway\nTo reduce the probability of sample condensation on the inlet pathway, the inlet was made as short as possible, was built with glass tube, and was provided with a controlled heating system that fixes its temperature at 100 \u00b0C. This temperature can be easily changed by an end user, thanks to the touch screen display interface mounted on the front panel of the sub-module. \nThe sensors temperature is also controlled by means of a thermoelectric module, underneath the sensor chamber, controlled by the main microcontroller. \nThe temperature and humidity level of the biosensors and the airflow passing through the sensor chamber are constantly monitored by means of two sensors collocated at the output of the sensor chamber.\nPump 1 fixes the airflow rate passing through the sensor chamber. Its velocity is controlled by the microcontroller and can be easily changed by an end user. \nMain electronic board\nA custom-made master board based on the ARM microcontroller technology is used to run D4. It is similar to that developed for V1 and consent to:\n\u2022\tDrive the 2 pumps; \n\u2022\tControl the 2-way valve;\n\u2022\tManage a SD memory card to save sensor data;\n\u2022\tManage 2 FRam memories;\n\u2022\tControl the inlet heating system; \n\u2022\tControl the sensor thermoelectric module by means of RS232 communication bus;\n\u2022\tCommunicate with the sensor board by means of an I2C bus;\n\u2022\tDrive the samplers and the pre-concentrator;\n\u2022\tImplement a user interface by means of a touch screen display driven by a SPI bus;\n\u2022\tImplement several communication protocols (USB, RS232; Bluetooth, SPI, Ethernet);\n\u2022\tRun the detector system through the different measurement phases;\n\u2022\tLaunch the multi-parametric software for sensor data recognition.\nSensor array\nThe sensor array comprises 8 diamond biosensors mounted into a circular sensor chamber. The biosensors were developed within WP5 and WP6. A custom-made sensor electronic board was developed within WP7. It allows sensor acquisition, calibration and self-diagnosis. \nSignificant results: \n\u2022\tIndividual submodule tested: electronics and software validated;\n\u2022\tLow level software \u2013 written and tested \u2013 activation and scanning software for the microcantilevers tested and functional.\n\u2022\tSensor interface electronics and software designed, constructed and tested;\n\u2022\tMaster controller interface to WP2 functional were verified;\n\u2022\tFirmware for running the sensor sub-module was written and tested;\n\u2022\tParametric pattern recognition software written, tested and implemented in the master controller of WP2;\n\u2022\tV2 built and tested in field trials in Athens airport in March 2015\nDeviations: \nDescription of problem \/ deviation\tSuggested \/ undertaken corrective action\tImpact on other WPs, budget\/PM plan or work plan\nDelays in testing individual components \u2013 redesign necessary in some cases\tCorrective action taken in all cases\tSlight delay possible in delivery of V1 instrument\nCondensation of the sample on the inlet pathway having the effect of reducing sensor sensibility, reproducibility and increases the limit of detection of the system.\tCorrective action taken - inlet pathway as short as possible, Inlet glass tube, heating of the inlet path \tNone\n\n\n\n3.5\tWP 5 Biosensor Development\nWP number\t5\tStart date or starting event:\tM1\nWP title\tBiosensor Development\nActivity type\tRTD\nParticipant ID\t1\t9\t\nPart. short name\tCEA\tESIEE\t\nPM total (for 60M)\t60\t42\t\nPM spent in period 1\t27,29\t27,84\t\nPM spent in period 2\t32.7\t27,82\t\nPM spent total\t60\t55.66\t\nOverview of the progress of the work during the period\nThe work carried out during this first period led to the following achievements:\n\u2022\tSuccessful immobilisation of OBPs and MUPs onto diamond surfaces \n\u2022\tOptimisation of diamond microcantilever transducers in terms and sensor actuation and read out (OK) \n\u2022\tTransfer of OBPs and MUPs immobilisation protocols onto sensor active surfaces\n\u2022\tValidation of the whole sensor principle\nWork progress in each task \nTask 5.1 V1 of Biosensors\nThe technologies developed within SNIFFER included some sampling as well as detection sub-systems. The scientific background of the sampling steps has already been described in deliverable D311 and the development of V1 sub-systems for the sampling and pre-concentration of target analytes were presented in DL314. \n\nWP5 has dealt with the development of single microcantilever diamond sensors combining the odorant proteins selective coatings. This work has involved the development of diamond microcantilever transducers with optimum performances in terms of mechanical dynamic properties, which are linked directly to the sensitivity of the sensors. This work has involved also the development of chemical grafting procedure in order to immobilize covalently the bioreceptors (MUPS and OBPs) onto the diamond surface of the transducers. The specifications of such sensors have been defined in D511. Those technologies were then exploited in WP4 for the sensor array development which takes also in consideration the problematic associated with multiplexing the cantilevers. \n\nThe tasks that have been involved in V1 of biosensors included:\nThe development of diamond microcantilever chips\nThe development of protein covalent immobilization procedure on diamond surfaces\nAssessment of the performance of the immobilized proteins using pre-existing surface acoustic wave available in the CEA laboratory.\nTransfer of the immobilization procedures onto actual diamond microcantilever transducers\nDevelopment of a gas cell that can accommodate several sensors together in view of their integration into a sensor array. \n\nTask 5.2 V2 of Biosensors\nDuring the first 18 months of the project a prototype V1 of biosensor has been specified and developed. The specifications of such sensors have been reported in DL5.1.1 and the results from the work and validation of V1 described in details in DL5.1.2. Furthermore, specifications of the version V2 of biosensors have been reported in DL5.2.1 which covers all aspects of biosensor development including the transducers, the biological sensitive coatings immobilization (OBPs and MUPs) as well as the multisensor array gas cell. \n\nIn the version V2, the main improvements on the microcantilever are as follows. Firstly the read out metal gages were replaced by polysilicon gages featuring much better gage factor. Secondly, the metal used for contacting to the sensors was changed to insure better resilience during the protein immobilisation process. Finally, the cantilever geometrical parameters were modified in order to improve sensitivity. Two different approaches have been considered in V2: bulk diamond cantilevers and diamond coated silicon cantilevers.\n\nAn efficient protocol to immobilise both MUPs and OBPs covalently onto the diamond surface of the cantilevers was proposed with V1. This protocol proved to be highly efficient therefore it was also used in v2 to immobilised new proteins on diamond sensor . \n\nA V1 gas cell was designed according to the SNIFFER project specifications of biosensor development. In collaboration with ESIEE-Paris the sensors in the cell could be interfaced electronically to the acquisition board to display and validate that the complete sensor array system was operational.  The development of the V2 multi-sensor gas cell was a follow up of the first cell that was developed in the first part of the SNIFFER project and patented. It was described in DL523. V2 multi-sensor gas cell takes into account the feedbacks received from the work experience and assessment gained with the first gas cell. However most of the improvements made were mostly targeting a reduction of fabrication cost since the performances of the first cell were already acceptable. \n\nFinally a range of new OBPs and MUPs have been provided by partners UNIMAN UNIPD and GTP that were characterised either using reference diamond SAW transducers or diamond cantilevers.  \n\n1.\tV2 of microcantilevers\n\n1.1.\tBulk diamond microcantilevers\n\nThe strategy elaborated to actuate and read-out the sensor signals was described in DL5.1.1. It consists of actuating all cantilevers in the gas cell array using a single external piezo-transducer located underneath the different sensor chips. The sensor signal is thus obtained from the measurement of the strain resistance of gages placed at the anchor on the cantilevers. The optimum gages parameters were simulated and reported in DL5.1.1. The following gage characteristics were obtained: gage factor \u003E 20 and resistance value \u003C 1 k\u2126. Neither metal nor boron doped diamond gages used in the V1 of cantilevers were found not to satisfy those conditions. Therefore it was decided for V2 to use polysilicon gages. Indeed polysilicon gages have proven to be efficient in the area of silicon MEMS with gage factors in the order of 100.\nHence the new strategy consists of embedding polysilicon gages into the diamond microstructures. Thus a process to include polysilicon piezoresistive gages into diamond micro cantilevers was developed as described in Fig.1. \n  \nProcess flow to the fabrication of diamond micro cantilevers including embedded polysilicon piezoresistive gages\n\nFor mass detection, we demonstrated that the higher the resonance frequency of the cantilever, the higher the detection sensitivity. However, high resonance frequency cantilevers are geometrically short and mechanically rigid cantilevers therefore reducing the electrical signal at the output of the piezo-resistive strain gages. The main drawback with this design is that the level of signal delivered by the gages is also low. Even though with the electronic amplification it is possible to follow the frequency resonance of the cantilever, the signal strength is of the same order of magnitude as the noise level. To improve the electrical S\/N at the output of the gages, it was decided to explore 5 different lengths of cantilever. The dimensions and theoretical resonance frequency (first mode) of each design is indicated in Table 1. Fig. 2 gives a graphical representation of 5 cantilever designs. Additionally, Fig. 3 presents the resonance frequency variation as a function of cantilever length and thickness.\n\nDimension and theory frequencies of new cantilevers design\nDesign\tCantilever characteristic\n(length \/ width)\tGage dimensions on the cantilevers (length \/ width )\tTheoretical Frequency(Hz) \n\t\t\te=1 \u00b5m\te=3 \u00b5m\n1\t210 \u00b5m \/140 \u00b5m\t150 \u00b5m \/ 10 \u00b5m\t61700\t185300\n2\t310 \u00b5m \/140 \u00b5m\t225 \u00b5m \/ 10 \u00b5m\t28300\t85000\n3\t410 \u00b5m \/ 140 \u00b5m\t300 \u00b5m \/ 10 \u00b5m\t16200\t48600\n4\t510 \u00b5m \/140 \u00b5m\t375 \u00b5m \/ 10 \u00b5m\t10400\t31400\n5\t610 \u00b5m \/140 \u00b5m\t450 \u00b5m \/ 10 \u00b5m\t7300\t22000\n\n\n \nFabrication masks design for the five cantilever lengths\n\n\n \nResonance frequency as a function of Cantilever length and thickness\n\n\nDiamond microcantilevers were fabricated according to the above descriptions. Figure 4 shows an optical microscope image of a cantilever during the fabrication process.\n\n \nOptical microscope image of a polysilicon gage during the fabrication process in step 7\n\n\n1.2.\tDiamond coated Silicon microcantilevers\n\nCEA-LCD developed a specific process to graft OBPs and MUPs on diamond surfaces. The process was validated by chemical surface analysis (XPS, EIS, fluorescence imaging) and for sensing performances using diamond coated SAW devices. Diamond coated SAW sensor is a technology that was developed prior to the Sniffer project by CEA-LIST.  Here the process involves spreading diamond nanoparticles over the sensor substrate and then fix them by a short CVD plasma growth in order to obtain at the end a thin polycristallin diamond coating over the surface. \nBased on the same approach, silicon cantilevers were fabricated with embedded polysilicon gages and coated with diamond. This approach enables high control of the whole fabrication process using previously developed processes on silicon while still offering all the advantages of the diamond coating for chemical grafting and stability.  Hence the two approaches (bulk diamond and diamond coating) will be later assessed in terms of sensing performances. \nThe silicon cantilever process uses a specific silicon substrate call SOI: Silicon-On-Insulator. These substrates are composed by 3 layers:  a first silicon layer (Device), a silicon dioxide layer, and a second silicon layer (Handle). When compared to standard plain silicon wafers, SOI allows high accuracy control over the cantilever thickness and the integrated silicon dioxide layer is naturally acting as a stop layer during backside etching and release of the resonant structure. \nThe process to manufacture silicon cantilever using SOI wafer is summarised in Fig.5.\n\n         \nDescription of process for Silicon cantilevers\n\nSimulations were performed to determine the thickness of the silicon cantilevers leading to resonance frequency within the same range as the diamond cantilever already used in the project. As shown in Fig. 6, a silicon layer of 6.5\u00b5m thickness is adapted to the resonance frequency range targeted by this project.\n\n \nSilicon cantilever resonance frequency vs. thickness \n\nSilicon cantilevers were fabricated using the SOI process described in figure 5. A Micro Scanning Laser Vibrometer was used for characterizing the frequency response of the designed cantilevers. This equipment consists of a laser source emitting between 620 nm and 690 nm, an optical fibre interferometer OFV511, a demodulator OFV3001 (Polytec GmbH) and a network analyzer Agilent 89410A. Figure 7 shows the frequency measured for each cantilever of different length (represented by points) and the computed theoretical curve of frequencies (represented by a black line). The results show that the measured frequencies match well the expected values from calculation. \n\n\n \nResonance frequency measured (points) and computed (black line) as a function of Cantilever length\n\nCantilevers were then electrically connected in order to measure the S\/N ratio out of the strain gages using the read-out electronic developed for V1 Sniffer device.   Fig. 8 shows the typical response for a V1 and a V2 cantilevers with the same length (here 210 \u00b5m). Here the S\/N ratio was improved by 25% hence leading to better sensitivity of the resulting sensors. Example of spectral responses are summarised in Table 2 for V1 and V2 cantilevers, again confirming those results.\n\n \nExample of response measured using SNIFFER V1 read out electronic for a V1 and a V2 cantilever with the same length 210 \u00b5m\n\nExamples of spectral responses for both V1 and V2 cantilevers  \nChannel\tID\tLength (um)\tFreq (KHz)\tS\/N\tResistance R1 (cantilever) and R2 (fixe) (K\u2126) \tSpectrum\n1\tSOI 2-6A515\n(V2)\t610\t20,700\t40.0\tR1 = 61,59\nR2 =62,85\t \n3\tSOI1-2B721\n(V2)\t210\t131,770\t40.0\tR1 = 26,55\nR2 =27,84\t \n4\tDiamond Q24\n(V1)\t210\t63,870\t29.75\tR1 = 23,16\nR2 =22,29\n\t \n\nFinally the metal contact used in V1 proved not to be performing well. This was due to the lack of robustness of the Cr\/AU layers when the cantilevers were placed in the diamond growth reactor to undertake plasma treatment in view of hydrogen terminate the surface, a step that is necessary in order to later graft the ligand binding proteins. Hence those contacts were replaced in the V2 by Ti\/Pt contacts that were found to withstand perfectly the same plasma conditions. Fig. 9 shows comparative images of V1 and V2 generation cantilevers.\n\n \nMicroscope image of two cantilevers, V1 and V2, summarizing all improvement implemented in V2 when compared to V1.\n\n2.\tV2 of multisensory gas cell\nThe development of the V2 multi-sensor gas cell is a follow up of the first cell that was developed in the first part of the SNIFFER project and patented. It was described in DL512. V2 multi-sensor gas cell takes into account the feedbacks received from the work experience and assessment gained with the first gas cell. The improvements that have been made are still in line with the claims made in the first patent; therefore this new cell is also protected by the same patent. A schematic view of this new cell is shown in Fig.10 (left). As a reminder and for comparison purpose a view of the former V1 gas cell is also shown (Fig.10 right). These schematics show clearly that V2 of the gas cell has been simplified when compared to the V1. \n                               \n\nSchematic view of - Left: the V2 multi-sensor gas cell, - right: the V1 multi-sensor gas cell\n\nSeveral changes have been made over version V1 in particular to improve both cost efficiency and performances. In particular: \n\u2022\tThe 2-parts base of the device (holder + embodiment) has been replaced by a single part (in grey colour in the V2 schematic). Since this part of the device does not see gas the material has been replaced from stainless steel to aluminium for cost efficiency. \n\u2022\tThe cantilever holder (green part on both figures) has been simplified in version 2. Indeed in version 1 the inside of this part had to be micro-machined in order to be able to position the sensors with high accuracy so as to align the 24 contacting pin with the metal contact pads of the 8 cantilevers. This operation contributed to increasing very significantly the fabrication cost. Therefore in the version2 of the cell, micro-machining of this part was not done. Instead a new part was fabricated from a sheet of stainless steel that was laser cut. A photograph of this new part is shown in the following figure.\n \nView of the part designed to hold and align the cantilevers inside the gas cell\n\nFinally, the lid of version 1 which was also made of two parts has been simplified to one part. This facilitates significantly the replacement the opening\/closing of the cell and therefore the replacement of sensors. The design of this lid was also modified to improve sealing of the cell to ensure a perfectly gas tight chamber. \n\nThe above modifications allowed reducing the fabrication cost by typically a factor 4 (for small batches processing) and reducing very significantly the size and weight of the overall cell. A photograph of the finished cell is shown in the following figure.\n\n \nPhotograph of the V2 of multi-sensor gas cell\n\n3.\tSelection of proteins as selective coatings\n\nBased on previous results obtained by the UNIMAN and UNIPD on one side while performing competitive binding experiments in the liquid phase, and on the results from the first series of laboratory trials, a shortlist of LBPs has been drawn that will be used to perform the final testing. These proteins are listed in the following table.\n\n\nList of OBPs and MUPs to be used for the final lab trials and field trials\nOBPs\tMUPs\nGTP_pigWT\tMUP3 (wt)\nGTP_pigF88W\tMUP20 (wt)\nGTP_AgamOBP1\tMUP20L88Q\nGTP_AgamOBP4\tMUP20L124V\nGTP_AgamOBP5\tMUP20Y103R\nGTP_AgamOBP19\tMUP20Y103D\nGTP_agamOBP47\tMUP7\nGTP_AgamOBP1_S82P\tMUP20N107L\n\nGTP_AgamOBP1_K93H\t\nGTP_LmigOBP1\t\n\nThe proteins will be grafted onto the diamond sensors using the chemical protocol developed in version 1 of sensors consisting of grafting a NTA-Ni radical onto the diamond surface followed by chelating to the Nickel via the 6His tags of the proteins. This approached was found to give the best results in terms of detection. When wild type proteins with no 6His tag will be used the other approach involving attachment of hexanoic acid radical followed by EDC\/NHS coupling will be used, as described previously. \n\n4.\tSensor integration into the measurement gas cell\n\nThe transducers developed of the V2 of biosensors have been described specifically in DL5.2.2. Their description details and fabrication steps can be found in this deliverable. Fig.13 shows a picture of eight diamond micro-cantilevers placed in the gas cell. \n \nView of a V2 micro-cantilever placed in the gas cell\n\nIn order to validate the V2 of biosensors, the cantilevers were characterized in the cell gas and configured to work on the standard test conditions. The performances of the system were assessed using different cantilever lengths. A micro-scanning laser vibrometer (Polytec OVF511 and OVF3001) and a spectrum analyzer (Tektronix RSA3303B) were used to characterize each sensor in terms of resonance properties and the data were compared to the resonance profiles extracted by our autonomous SNIFFER system. Frequency profiles were measured and normalized within a bandwidth of 3 kHz around each resonance frequency. An example of comparison between the two methods is depicted on Fig. 14. Table 4 presents the experimental figures measured to compare the optical reference data with the electrical measurement with the SNIFFER readout system. \n\n \nFrequency Response of eight cantilever of 510um length.\n\nComparison of the optical reference measurements with the electrical measurements\nSensor\nnumber\tCharacteristics\t\tSensor number \tCharacteristics \n\t\u0394 Frequency(Hz)\t\u0394 Q. Factor\tRMSE\t\t\t\u0394 Frequency(Hz)\t\u0394 Q. Factor\tRMSE\n1\t26.31\t101.75\t0.108\t\t5\t22.17\t44.93\t0.154\n2\t6.61\t100.22\t0.091\t\t6\t0.50\t8.28\t0.090\n3\t9.72\t77,89\t0.110\t\t7\t10.00\t52.27\t0.111\n4\t12.31\t37.82\t0.057\t\t8\t2.59\t48.56\t0.098\nNote: Root mean square error (RMSE) is evaluated using the difference between the two frequency responses corresponding to the two measurement methods presented in this work (evaluated using the normalised magnitude above -8dB of the centre frequency).\n\nOptical and electrical measurements were also carried out for each cantilever placed in the gas cell to analyse the influence of the relative location on the piezoelectric cell on the measured resonance frequency and quality factor. The results are presented in the following figure.\n\n  Resonance Frequency and Quality Factor of a V2 micro-cantilever placed in different positions in the gas cell\n\nFrom these preliminary tests we were able to quantify the influence of the closed gas cell and the electronic circuit on the resonance frequency measurements. Further experiments were performed in order to validate the operation of the sensors and to demonstrate the performance of the whole system, including gas cell and electronics, to detect sensors frequency shifts.\n\nA set of eight silicon cantilevers of 410\u00b5m length were selected to be tested. The below table lists the sensors characteristics, as well as the post-treatments received by each sensor, respectively.\n\nSensors characteristics and treatment in experiment 1\nCantilever\tUnique code\tResonant Frequency(KHz)\tTreatment\tCantilever\tUnique code\tResonant Frequency (kHz)\tTreatment\n1\t4B118\t54.368\tDiamond nanoseeded\t5\t4B113\t49.353\tBare cantilever\n2\t4B23\t58.981\tDiamond nanoseeded\t6\t4B38\t48.234\tBare cantilever\n3\t4B318\t47.304\tDiamond nanoseeded\t7\t4B813\t52.646\tBare cantilever\n4\t4B218\t49.742\tDiamond nanoseeded\t8\t4B58\t47.410\tBare cantilever\n\n\nIn order to verify the influence of the humidity on the sensors, the gas (nitrogen) flow in the cell was set to 500 sccm\/min. After stabilization of the system (regarding temperature, flow, etc.) we proceed to the injection of humid air at 100%. The results of this experiment are presented in figure 16.\n\n \nFrequency shift of sensors subjected to cycles of dry air and moist air\n\nIn this experiment, sensor 7 and 8 could not be electronically compensated and the output signal was discarded in this analysis. In the first graph of figure 17 we can identify a frequency shift of some tens of Hertz for the nanoseeded cantilevers. This demonstrates that the frequency shift depends on the surface conditions of the cantilevers.\n\nIn the following experiment, using the same set of eight cantilevers described in table 5, cantilever 5 and 6 were coated with PMMA polymer in order to obtain sensitivity to chemicals. The characteristics of sensors are listed in thetable below.\n\nSensors characteristics and treatment in experiment 2\nCantilever\tUnique code\tResonant Frequency(KHz)\tTreatment\tCantilever\tUnique code\tResonant Frequency (kHz)\tTreatment\n1\t4B118\t54.368\tDiamond nanoseeded\t5\t4B113\t49.580\tPMMA\n2\t4B23\t58.981\tDiamond nanoseeded\t6\t4B38\t48.335\tPMMA\n3\t4B318\t47.304\tDiamond nanoseeded\t7\t4B813\t52.646\tBare cantilever\n4\t4B218\t49.742\tDiamond nanoseeded\t8\t4B58\t47.410\tBare cantilever\n\nAfter being polymer-coated, cantilever 5 and 6 have changed their resonance frequency due to the added mass. Before injecting solvent vapours on the gas cell, the gas flow was set at 500sccm\/min (nitrogen) and we waited for two minutes in order to stabilize the measurement. The chemicals used in this experiment are: 1-octanol, acetone, ethanol, DNT and Nitromethane (a precursor used to prepare nitroaromatic explosives). The resulting sensor responses during this experiment are presented below. \n\n \nFrequency shift of sensors subjected to cycles of different chemicals\n\nFrom the results presented in figure 17, we can clearly observe that sensors 5 and 6 have shown the best responses to ethanol, acetone and nitromethane. For the detection of ethanol, the frequency shift can reach up to 1 KHz and for nitromethene we can observe several hundred of Hz of variation. Bare cantilevers (7 and 8) respond only to ethanol while the nanoseeded cantilevers (1 to 4) respond to acetone and ethanol with a frequency shift of about 40Hz.\nImprovements on the functionalization steps can be done to achieve higher sensitivity and selectivity. Further lab trials are to be carried out on diamond cantilevers as well as silicon cantilevers, using LBPs as sensitive coatings.\nSignificant results: \nTwo procedures have been developed and characterised in order to immobilise OBPs and MUPs onto diamond surfaces. The first procedure enables covalent attachment of the proteins with random orientation over the surface. The second approach allows orientation of the proteins over the surface through the use of hystidine tag. \nBoth protein attachment procedures were transferred successfully to diamond microcantilevers. The second approach appeared to give significantly better sensing performances and will be chosen for the V1 prototype.\nA gas cell was designed according to the SNIFFER project needs. The cell can accommodate 8 sensors in parallel. In collaboration with ESIEE-Paris the sensors in the cell could be interfaced electronically to the acquisition board to show that the complete sensor array system is operational. \nSensor performance assessment was carried out with chemical simulant (DNT, 2-methylpyrazine). A test campaign with real drugs and explosive has been carried out in July at CREAL (France) to validate the sensors with substances of interest for the SNIFFER applications.\n\nA set of improvements has been implemented in version 2 of cantilevers. They can be summarised as follows: \n\u2022\tChange of piezo gage material from BDD to polysilicon with much higher gage factor. A dedicated process had thus to be developed in order to embed such gages into bulk diamond cantilevers.\n\u2022\tChange of geometry of both the cantilevers and strain gages in order to increase S\/N ratio. This change was shown to pay off since and increase up to 25% in S\/N could be observed with the V2 cantilevers\n\u2022\tChange of metal contacts to increase robustness, in particular during the proteins immobilization process, and long term stability of the sensors. \nMoreover an alternative process based on the fabrication of Si cantilevers from SOI wafers and then coating of the cantilevers with diamond has been proposed. \nA new version of gas cell was also developed.. This new gas cell can accommodate up to 8 individual sensors that can be actuation simultaneously by a single piezo-actuator embedded into the gas cell. The gas cell has been simplified considerably over the first version in order to reduce fabrication cost and practicality of replacing the sensors in the cell.\n\u2022\tThe cantilevers were functionalised with various coatings in order to verify the chemical sensitivity of the complete sensing module. The sensors were sensitive to various vapours including acetone, ethanol or nitromethane, a precursor for many explosives.\nDeviations: None\n\n3.6\tWP 6 Protein Engineering\nWP number\t6\tStart date or starting event:\tM1\nWP title\tProtein Engineering\nActivity type\tRTD\nParticipant ID\t2\t8\t10\t\nPart. short name\tUNIMAN\tUNIPD\tGTP\t\nPM total (for 60M)\t15\t36\t12\t\nPM spent in period 1\t9\t22\t8,48\t\nPM spent in period 2\t23\t28\t7,64\t\nPM spent total\t32\t50\t16,12\t\nOverview of the progress of the work during the period\nThe major objectives were :\n\u2022\tSelection of LBPs\n\u2022\tTailoring of protein recognition site conformation for a selected analytes through selective mutagenesis\n\u2022\tProduction of proteins by expression and the industrial scale-up of LBP productions\n\u2022\tImmobilisation of proteins on to diamond coated cantilever substrates\n\u2022\tTesting of sensitivity and selectivity of immobilised proteins to selected analytes\n\u2022\tValidation of stability, repeatability and influence of environmental factors.\n\nTask 6.1 \u2013 V1 of proteins \nTask 6.1 concerned selection of suitable proteins for Version 1 of the SNIFFER device. Close interactions between UNIMAN, UNIPD, GTP as well as the user partners CEA and ESIEE were established to achieve the goals. This involved selection of LBPs, followed by in-house protein expression, then industrial scale production by P10 (GTP), then testing the binding properties of the produced proteins towards the targeted analytes. Initial experiments were performed by UNIMAN and UNIPD aimed at carrying out TASK 6.1 \u2018V1 of Proteins\u2019. A first batch of OBPs was produced by UNIMAN using a bacterial expression system. It was composed of four different OBPs. Moreover, a functional recombinant MUP (rMUP) was expressed by UNIPD. Affinities and selectivities of the selected OBPs and the rMUP for target analytes were investigated. Processes were developed for producing and purifying the OBPs and MUPS using appropriate molecular biology technique, to produce test batches in small quantities. Fluorescent binding experiments were performed to test the affinity of the mutant pOBP-FW and the rMUP to target explosives and drugs. The gene sequences of selected proteins were transferred to partner GTP, who commenced the process of scale up for large scale production of the ligand binding proteins.\n\nTask 6.2 \u2013 V2 of proteins \nTask 6.2 was concerned with the preparation for V2 of proteins (OBPs and MUPs) for the second stage of the project. This involved the design of methodology for cloning, expressing and modification of the selected proteins, as well as the quality control criteria for assessing the viability of the produced proteins. It involved the definition of the boundaries of acceptable performance taking into account dynamic range, affinity, selectivity etc. In conjunction with partner GTP, compatibility with industrial production methods and application-focused needs such as response time, robustness and influence of environmental variables were optimised. \nWork progress in each task \nTask 6.1 \u2013 V1 of proteins\nThis involved close interaction between UNIMAN, UNIPD, GTP as well as the user partners CEA and ESIEE. Consultant Prof. Paolo Pelosi from University of Pisa and his laboratory were heavily involved in the initial stages and provided genetic material to UNIMAN and practical help to both UNIPD and UNIMAN. This involved selection of LBPs, followed by in-house protein expression, then industrial scale production by P10 (GTP), then testing the binding properties of the produced proteins towards the targeted analytes.\nInitial experiments were performed by UNIMAN and UNIPD.\nThis task included:\n\u2022\tSetting up recombinant protein expression at UNIPD and UNIMAN\n\u2022\trOBP and rMUP expression and purification at UNIMAN and UNIPD\n\u2022\tEvaluation of rMUP ligands by GC\/MS - UNIPD\n\u2022\trMUP binding assay in two independent Labs - (UNIPD and Prof. Pelosi Lab in Pisa)\n\u2022\tIn silico screening of natural  MUP mutants - UNIPD\n\u2022\tSelection of alternative natural MUP isoforms  - UNIPD\n\u2022\tSetting up gene sequences for OBPs and MUPs \u2013 UNIMAN, UNIPD\n\u2022\tEngineering of expression plasmids for OBPS and MUPs \u2013UNIMAN, UNIPD\n\u2022\tBacterial transformation and protein Expression of OBPs and MUPs \u2013UNIMAN, UNIPD\n\u2022\tCharacterisation of newly expressed recombinant proteins \u2013 UNIMAN, UNIPD\n\u2022\tBinding assays to characterise the proteins against target analytes, UNIMAN, UNIPD\n\nA first batch of OBPs was produced by UNIMAN. It was composed of four different OBPs (1PDOM from the insect wasp and F88W from the pig were characterised). Moreover, a functional recombinant MUP (rMUP 8) was expressed by UNIPD. Affinities and selectivities of the selected OBPs and the rMUP for target analytes were investigated. Processes were developed for producing and purifying the OBPs and MUPS using appropriate molecular biology technique, to produce test batches in small quantities. Fluorescent binding experiments were performed to test the affinity of the mutant pOBP-FW (UNIMAN) and the rMUP (UNIPD) to target explosives and drugs. Samples of proteins developed were sent to CEA for immobilisation on to diamond substrates and subsequent characterisation of immobilised protein. The gene sequences of selected proteins were transferred to partner GTP, who commenced the process of scale up for large scale production of the ligand binding proteins.\n\nMajor Urinary Proteins\nEngineered yeast (Pichia Pastoris) expressing recombinant mouse Major Urinary Protein (rMUP) was kindly provided by Dr. Elena Ferrari (University of Parma \u2013 Italy). Heterologous expression of rMUP was carried out according to a protocol previously described [Ferrari et al., 1997]. \nThe optimal volume of biomass for the best rate of protein production was determined and rMUP was expressed, purified and concentrated.  A first evaluation on possible endogenous ligands, naturally present in the purified rMUP, was performed by Solid Phase Micro Extraction (SPME), followed by GC\/MS analysis according to a previously described procedure [Redaelli et al., 2006]. The data were processed using the software MS data Review (Varian, Italy) in comparison with our in house spectra library and the NIST Mass Spectra Database (version 06, NIH 2006). Binding assays against target analytes were performed in two independent facilities (UNIPD and Pisa Lab). \n[Ferrari E, Lodi T, Sorbi RT, Tirindelli R, Cavaggioni A, Spisni A. Expression of a lipocalin in Pichia pastoris: secretion, purification and binding activity of a recombinant mouse major urinary protein. FEBS Lett. 1997. 401(1):73-77.]\n\nTask 6.2 \u2013 V2 of proteins\nThis involved close interaction between UNIMAN, UNIPD, GTP as well as the user partners CEA and ESIEE. Work done in this task has been reported in detail in DL6.21, DL6.22 and DL6.23.\n\u2022\tA number of OBPs and MUPS mutants were selected through in silico screening towards target analytes via ligand docking experiments and subsequent data analysis.\n\u2022\tMethods of protein expression were optimised and selected mutants of OBPs and MUPs were expressed, and their affinities to target analytes tested, together with stability and repeatability of performance in detecting and recognising sniffer target analytes. \n\u2022\tQuality control considerations have been analysed and steps taken to ensure the performance of manufactured proteins.\n\u2022\tThe scale-up of protein production was carried out by partner GTP.\n\u2022\tCharacterisation of the GTP produced proteins (both WTs and mutants) in terms of their affinities, and stability of performance in detecting and recognising sniffer target analytes was carried out at UNIMAN and UNIPD.\n\u2022\tProteins were sent to Partners ESIEE and CEA to be immobilised onto the transducers.\n\u2022\tThe characterisation of the proteins once immobilised onto the transducers was carried out in partner (CE and CREAL) and that have the appropriate licenses for target compounds. \nSignificant results: \nSuccessful site directed mutagenesis techniques have been developed by UNIMAN and UNIPD involving in silico modelling of the binding sites and docking screening approaches. The approach has been used to design mutant OBPs and MUPs leading to new proteins with enhanced binding properties towards Sniffer target analytes. \nEleven 11 OBPs (eight Wild Type proteins and three mutants were produced in industrial scale by GTP (Partner 10) and then characterised; these proteins were used to develop SNIFFER devices.\nTen MUPs (three WTs and seven mutants) were produced by UNIPD and GTP and then characterised; these proteins were used to develop SNIFFER devices.\nThe scale-up of protein production by partner GTP means that OBPs and MUPs can be produced in 100mg to gram quantities. This is a significant technical success that would not have been possible without the SNIFFER project to drive this development.\nMajor result:\nWe consider that the development of new and novel Ligand Binding Proteins for the SNIFFER project has been an extremely successful part of the project and this has provoked innovative and novel technical advances that have also generated significant new intellectual property applicable to the wider biosensor field apart from the objectives of the SNIFFER project. The incorporation of these proteins into arrays has never been achieved before. The characterisation of the proteins once immobilised onto the transducers carried out in partner laboratories (CE and CREAL), indicate good sensitivities to target analytes.  The recent field trial carried out in Athens has confirmed that the LBPs developed in this project are indeed performing well in the recognition of the sniffer target analytes. \nDeviations: None\n3.7\tWP 7 Self-Diagnostic Capabilities\nWP number\t7\tStart date or starting event:\tM1\nWP title\tSelf-Diagnostic Capabilities\nActivity type\tRTD\nParticipant ID\t2\t9\t\nPart. short name\tUNIMAN\tESIEE\t\nPM total (for 60M)\t6\t18\t\nPM spent in period 1\t2\t13,35\t\nPM spent in period 2\t0,3\t3,50\t\nPM spent total\t2,3\t16,85\t\nOverview of the progress of the work during the period\nWP7 is completely dedicated to self-diagnosis and calibration. \n\u2022\tTask 7.1, Self-Diagnostic Sub-System, aims at the development of the electronics for performing the self-diagnosis. The objective  of the first period was to develop a prototype that can be interfaced with the gas cell to extract the resonance frequency of 4 cantilevers in real time. This working V1 prototype was then refined during the second period of the project toward an electronic that can address a full array of 8 diamond resonant sensors and bring full self-diagnostic and self-characterization feature. The objective of reaching precise resonant frequency tracking while providing the real-time information of the characteristics and health of the sensors has been successfully reached. Moreover self-characterization information are used to calibrate the electronic and the digital signal processing algorithm developed to improve overall system sensibility and robustness. .\n\u2022\tTask 7.2, Reference Samples for Calibration, is involved in the definition of the calibration procedure and reference samples to be used for the calibration procedure.  The objective of this task are to define the calibration process for the prototype as well as the reference samples and the rationale that led to their selection.\nTask 7.1 \u2013 Self diagnostic sub-system\n5.\tSensor mixed signal sensor read-out architecture Architecture\nWithin the SNIFFER project, WP7 is mainly concerned with the electrical interface and the acquisition of the resonance frequency of each sensor of the array. It also checks that the sensor array is working in the best possible conditions by mean of self-diagnostic and calibration processes. Those tasks have been integrated in an electronic sub-module that is interfaced on one side to the sensor array and on the other side to the global system controller (or supervisor).\nThe architecture that has been defined will provide the 3 main following functionalities :\n6.\tReadout: Design of electronic interface that provide an accurate readout of the useful values from each cantilever of the sensor array. The parameters that are measured are the resonance frequency of each cantilever as well as the relative amplitude at resonance. Information on each sensor of the array is communicated to the supervisor system. \n7.\tSelf-Diagnostic: Automatically characterization of each sensor of the array by getting its full spectral response to a monotonic harmonic excitation at different moments of the fabrication and during the life-cycle of the sensor. This is to detect any fabrication fault and monitor the sensor\u2019s behaviour during time with the perspective to inform the user that the sensor is not usable anymore. Furthermore the electrical characteristics of the sensor are also measured so that they can be compensated in the calibration sequence. This information is also indicative of the aging of the sensors.\n8.\tCalibration: Using the information from the self-diagnostic function to calibrate the electronic interface to get the best possible (highest SNR) readout from the sensor\nThese goals can be related to each other as shown in Error! Reference source not found.. It is expected from this process to calibrate the whole system to get the best possible readout by using the information gathered during self-diagnostic.\n\n \nInteraction between the 3 main functionalities of the sensor electronic sub-system\n\nThe block diagram presented shows the general architecture used to implement the sensor analog front end architecture :\n \nGeneral architecture of the electronic sensor interface sub-module\n\n\nEach cantilever of the sensor array is connected to a Wheatstone bridge structure that is polarized using a dedicated low noise polarization circuit to insure maximum signal to noise (SNR) ratio. The output of the Wheatstone bridge is amplified by a low noise instrumentation amplifier (1.5 nV\/\u221aHz at the operating frequency range and gain) before being filtered to focus on the frequency band of interest. \nOptional analog processing can also be implemented to directly extract information in the analog domain. This might prove to be useful if the digital counterpart of such processing is too expensive in terms of size, power consumption or development time.\nThis signal is digitized by an Analog to Digital Converter before reaching the digital processing unit. The digital processing unit is a Cortex M3 processor implemented in a LPC1768 microcontroller from NXP. A coprocessor unit based on Cortex M4 with floating point calculation capability has also been successfully interface with the sub-system. It is programmed to fulfil the functionalities described previously. It also creates a communication link with the supervisor system to receive instructions and provide the resonance frequency of each sensor.\nThe cantilevers are excited all together by a single piezo-electrical acoustic cell driven by a high current amplifier stage. The excitation signal is directly generated by the Direct Digital Synthesiser (DDS) controlled by the digital processing unit and then low-pass filtered.\nThe digital processing unit is be able to control parameters from the Analog Front End (AFE). By this mean it will individually optimize the acquisition of the resonance frequency of each cantilever.\n\n9.\tV2 Sensor electronic interface prototype design and validation\nThe generic architecture described in the previous section has be declined in 2 versions: V1 for the first period of the project and V2 for the second period. The V1 prototype of the subsystem has brought the proof of concept of the ability of this architecture to correctly address an array of diamond resonant sensors. Considering the challenges of the project and to reduce design risks, it has been decided to work iteratively on successive electronic prototypes. Each of them have implemented and validated each functionality of the sub-module. The V1 prototype architecture is shown on Error! Reference source not found.. It was only able to work with an array of 4 sensors and all the trimming and calibration steps were done manually. The cantilever-based sensors are inserted in a Wheastone bridge structure with a very low noise polarization circuit. The difference at the output of the bridge is amplified by a first low noise amplification stage with a static gain of 2000 due to very low signal level (\u00b5V range). This is followed by a second variable gain amplification aimed to trim the signal amplitude to a level useable by the rest of the acquisition chain.  The signal from each sensor is then bandpass filtered, its RMS value extracted and digitized before being processed by the NXP LPC1768 microcontroller. This microcontroller is also controlling the DDS that generates the signal used to drive the piezo acoustic cell below the gas cell. Separating the V1 architecture in three boards is aimed to simplify the validation of each functional block and to improve reusability of the boards over the whole project. This approach proved to be inefficient as assembling the 3 boards within the full integrated device on top of the gas analysis cell was difficult. V1 subsystem was also prone to power supply noise injection form the rest of the system into the low noise analog part of the circuit.\n\n\n  \nElectronic architecture and PCB implementation developed for V1 prototype of sensor electronic sub-module\n\nThe electronic circuit that has been developed and realized for the V2 prototype of the SNIFFER project improves the readout capabilities of the V1 prototype while introducing diagnostic and characterization elements. The mechanical partitioning of the sub-system has also been optimized to allow both easy access to the sensor and efficient integration in the global system. The global architecture of the system is depicted on Error! Reference source not found.. The system is divided in 2 boards:\n10.\tBoard B21 is the pre-amplification board which is located as close to the sensors as possible on the gas analysis cell. It is the analog front end for the sensor and it features digital calibration of the Wheatstone bridges. It is connected to the board B22 using 8 coaxial cables and a digital calibration control cable.\n11.\tBoard B22 holds the specific analog processing, the digital processing, calibration elements as well as the power supplies. The micro-controller on this board is not only performing and synchronizing the read-out of the resonance frequency of each of the 8 cantilever sensors but it is also digitally controlling the diagnostic and calibration elements of the whole sub-system. Moreover, it is also managing the serial communication link with the master micro-controller.\n \nGlobal architecture of V2 sensor sub-system\nParticular care has been given to the mechanical design of the sensor subsystem by taking into account the feedback from the users of the V1 prototype. The following guidelines have been defined for the V2 design:\n\u2022\tReduce as much as possible the size of the pre-amplification board (B21) in order not to interfere with the fluidic connection of the gas analysis cell.\n\u2022\tAvoid using a mechanically fixed connector between the pre-amplification board (B21) and the signal processing board (B22). This is to allow the sensors to be changed without major disassembly work.\n\u2022\tOversize the heat dissipation elements of the power supply circuit to cope with the system internal temperature.\nThe assembly of the gas cell, contact board (B20), pre-amplification board (B21) and signal processing board (B22) is shown on Error! Reference source not found..\n \nPicture of the V2 sub-system\n\n12.\tDigital signal processing algorithm for resonant sensor frequency detection\nOn the V1 prototype of the sub-system, the detection of the resonance frequency was done by detection of the maximum of the extracted profile. This method has proven to be working but the noise level on the frequency shift detection (at baseline) was between 10Hz to 30Hz depending on the quality of the sensor response.  In the second period, beside from the optimization of the hardware, we have focused on developing a digital signal processing algorithm based on function fitting to filter and optimize the frequency shift detection. The solution that has been chosen is based on fitting the frequency profile curve to a Cauchy function. The approach has been validated theoretically and implemented in real time on a Cortex M4 co-processor. Error! Reference source not found. shows the noise level on the frequency shift can be brought below 1Hz using this method. \n \nComparison of frequency shift error for a cantilever sensor resonance frequency shifting from 36,76kHz to 36,95kHz.\n\nMore practically, this digital signal processing sequence bring to the system the ability to exploit efficiently signals from sensors with a signal to noise as low as 5dB. This a tremendous 22dB gain compared to V1. \nFrom a system functionality point a view, the whole V2 sub-system is able to reliably follow the resonance frequency of up to 8 sensors at a time with a resolution of a 1Hz  and with 1 reading a second. An example of sensor array response is shown on Error! Reference source not found..\n \nExample of sensor array response detection using V2 electronic sub-system\n\n13.\tSolutions for calibration and self-diagnostic\nDuring the second period of the project, we introduced digitally controllable elements in the analog front end that brought the following functionalities:\n14.\tBridge compensation: Continuous balancing of the Wheatstone bridges. \nThis bring robustness to the measurement because the electrical equilibrium of the sensors is re-adjusted before every measurement (i.e. every second). The modifications of temperature or humidity in the gas analysis cell due to the sampling processing are not more corrupting the signal from the sensors. An example of this continuous calibration is shown on Error! Reference source not found..\n\n\n \nExample of cantilever array calibration voltages. Even if the bias voltage at the output of a sensor is drifting because of environmental parameters modification (T\u00b0, humidity, ...), sensor polarization is continuously re-calibrated\n\n15.\tAmplification gain setting: Continuous adjustment of the amplification stage for a 1V normalized signal.\n Each cantilever sensor in the array has a different electrical response because of the manufacturing dispersion. The amplitude of the sensor can also be modified by temperature changes or mechanical dispersion during the mounting\/unmounting of the array in the analysis chamber. By using a digitally controlled amplifier in the analog front end of the detection circuit, the amplitude of the signal at resonance of the sensor of the array can be normalized between 0,8 and 1V. The challenge in designing this feature lies in the introduction of digital control signal in the very sensible low noise analog front-end. \n\n \nFrequency profiles of the 8 resonant sensors of the array. Each sensors is individually amplified with a specific gain so that the global response of the array is normalized to between 0,8V and 1V.\n\n16.\tBandpass filter setting: Continuous adaptation of the low-pass filters to reject any noise above the resonance frequency\nEach of those calibration functionalities are possible because the characterization sequence occurring at system start up. \n\nTask 7.2 \u2013 Reference samples for calibration\nThe calibration process, as developed in SNIFFER, is an innovation in border security applications. The actual systems currently employed in this field often require to be handled by experts and it is practically impossible to understand, easily and on-line, if they are working properly and to resolve malfunctions.  \nThe calibration that has been defined will use a self-diagnostic module, implemented in the sniffers, to carry out an easy optimisation of the sensor performances in case degradation is detected. The evaluation of the sensor performances is   achieved by acquiring the full spectra response of the array of the reference samples and comparing them with the responses of the same substances acquired after grafting of the OBPs, stored in a database and used as sensor fingerprints.  Substances with similar molecular characteristics as the ones targeted by the V2 prototype are needed for the calibration but also need to be selected so that they are not dangerous to work with. \nSeveral substances have been selected as potential reference samples for the calibration of V2 devices, and are summarised in the table below.\n\nReference substance\tType\n2-nitro-p-cymene\tSimilar to  TNT\nMusk ketone\tSimilar to TNT\nPiperonal\tPrecursor of Amphetamines \nSafrole \tprecursor of MDMA\nProcaine \tsimilar structure to  cocaine\np-nitrophenol\todorant\nCarvone (-)\todorant\nCarvone (+)\todorant\n3-octanol\todorant \nIBMP\todorant\no-nitrophenol\todorant\n\nReference samples for the calibration of V2 devices\nSignificant results: \n\u2022\tDefinition of an evolutive architecture that lead to the V2 prototype with readout, characterization and calibration elements. It is a highly agile architecture that can be declined to other kind on resonant sensor and for a very large range of applications.\n\u2022\tThe challenge of provide a mechanical and electrical solution to access an array of 8 resonant sensors within a gas analysis cell has been successfully solved by the design of an innovative gas analysis cell with dedicated electrical interface.\n\u2022\tA robust, agile and fully functional prototype of the V2 prototype is now used for provide sensor data from sensors within a whole system designed for trace detection based on biomimetic odor detection. \n\u2022\tCharacterization and self-diagnostic hardware and procedure have been defined and implemented. They are the key elements in providing information on the reliability of the sensor array as well as for correcting in real time any drift in there operation.  \n\u2022\tFurthermore, a specific digital signal processing algorithm has been developed. It allows the systems to monitor the resonance frequency of each sensor of the array with a resolution of 1Hz while keeping the measurement rate up to 1 sample per second (8 resonant sensors data\/sec).\nDeviations: None\n\n3.8\tWP 8 Dissemination \u0026 Exploitation\nWP number\t8\tStart date or starting event:\tM1\nWP title\tDissemination \u0026 Exploitation \nActivity type\tOTHER\nParticipant ID\t1\t3\t12\t14\t\nPart. short name\tCEA\tST(SI)\u00b2\t3DSA\tARTTIC\t\nPM total (for 60M)\t2\t4,75\t3\t15\t\nPM spent in period 1\t1\t3\t0\t5,44\t\nPM spent in period 2\t1\t1,50\t3,10\t12,80\t\nPM spent total\t2\t4,50\t3,10\t18,24\t\nWork progress in each task \nTask 8.1 Dissemination activities\nThe SNIFFER partners took advantage of their network to spread news about the project, developed technologies as well as the European FP7 funding and collaboration frame. To support their actions, a few months after project launch, the project dissemination plan, from which all dissemination actions were implemented, including the communication strategy, was defined by all partners. \n\nThe SNIFFER Project dissemination strategy relied on diverse communication materials. It included the items listed below.\n\u2022\tThe project logo was drawn at earlier stages of the project. 4 options were presented to the partners who voted for the one to be reproduced on every SNIFFER document, including this deliverable (at the foot of the page). The project identity was then built around this logo, its spirit and colors. It resulted in the production of the project presentation \/ communication templates to be used by all partners when communicating about SNIFFER.\n\u2022\tThe project public website (www.sniffer-project.eu) which was launched to the public July 6th, 2012. \n\u2022\tThe project printed material covered a standard poster and leaflet. Their finalisation and printing were delayed due to the fact that the dissemination team waited for promising technical results to be included in their content. They were distributed to partners to support their speech at conferences at mid-project. \n\nDuring the course of the project and along project findings and promising results, partners made research developments and outcomes available to the scientific community through 6 scientific publications, mainly in peer-reviewed journals. The list of publications in available in the final reporting of the project.\nBesides the information given through the SNIFFER public website, the general public was also informed of SNIFFER outcomes and research domain through 2 press releases in widely spread journals.\n\nSpecific actions were taken to reach one specific target, the end-users, beyond task 8.3. During the last year of the project in May 2014, SNIFFER attended the FRONTEX European Day for Board Guard in Warsaw. This event was the opportunity to meet with FRONTEX national representatives, \u201cmember states\u201d, and to raise awareness among them of the SNIFFER developments and results. Beyond the networking opportunity with end-users, the ED4BG was also the opportunity to meet with other represented FP7-SEC projects. The final field trials in the airport environment were eventually the opportunity to further reach the end-users. They took place at Athens International Airport during March 2015 for 3 days. They revealed a genuine interest by airport authorities and the police for commercial devices based on SNIFFER technologies. \n\nAlso, SNIFFER partners were encouraged to participate in conferences and events related to the field covered by SNIFFER: specific technical bricks (e.g. proteins); artificial sniffing in general, border security, etc. At the end of the project, 16 of such participations have been counted all around Europe. Talks and lectures at national and international conferences were supported by the above described communication material.\nEventually, collaboration with other RTD projects was investigated from the start of the project. Early efforts paid off at project midterm when the SNIFFER Project built stronger links with 2 FP7-SEC  \u201ctwin\u201d projects (DOGGIES and SNIFFLES) who attended the first SNIFFER public event. Regular contact was made, as well as further networking for clustering with other RTD Projects. SNIFFER was represented at FRONTEX European Day for Border Guard in Warsaw in May 2014, which was a great opportunity to investigate further interactions with end-users and other projects. It resulted in strong links with FP7-SEC projects SNOOPY and HANDHOLD, which actively participated in the SNIFFER final public event. Furthermore, SNIFFER investigated other interactions with RTD projects such as OLAF Rats II, which deals with border security issues and illegal substances detection through rats.\n\nTask 8.2 Exploitations and IPR Management\nThe IPR team was assembled based on General Assembly members\u2019 votes. It never met as there was no IPR conflict. \nConcerning Exploitation, first discussions on the exploitation plan have begun at project mid-term and are reflected in D821. This initial exploitation plan was further discussed up to project end and let to promising exploitation perspectives. DL 822 reflects that, at project end, the exploitation activities related to SNIFFER results have already started or will do so in the near future. One patent has been issued 04 April 2014 under the reference \u201cFR1000235135\u201d related to Gas sensing cell with remote piezo- actuation of the cantilever sensors. Another patent related to Diamond MEMS with Polysilicon gauges is to be submitted in the near future. 9 major avenues for commercial exploitation are listed in the document as well as non-technical results.\n\nTask 8.3 User Advisory Group \nThe UAG was established at project start with the objective to gather experts coming from organisations involved in operational security interested in the SNIFFER technology. It first met in Manchester in July 2012. Some representatives from Athens airport (Greece) and the French customs were invited to give their input in the project. Each of them presented their organisation and the different devices they were used to develop for security on the airports (See DL 831). \n\nThe ST(SI)\u00b2 identified some new partners to join the UAG. SNIFFER made contact with Frontex (European agency dealing with security on borders) which is interested mainly by the detection of human beings (illegal migrants) but also the German police and the UK border unit.\n\nThe UAG was then understood in broad considerations: it did not only cover the UAG members stricto sensu (involved on a regular basis during the 3 years of the project) but also the overall group of end users that, somehow, got punctually involved in the SNIFFER Project by providing a specific input during an informal conversation, punctual help to cover specific project tasks, or simply followed the SNIFFER project through its public events.\nDuring year 3 of the project, another UAG-specific physical meeting did not seem bringing added-value to the project developments. Regular informal meetings, email and phone exchanges between the UAG (in its broad understanding) with some SNIFFER Partners as interfaces for the SNIFFER consortium proved sufficient. The SNIFFER end-user partners were greatly involved in the project. The Greek partner 3DSA, the French Ministry of Interior (ST(SI)\u00b2) and its police laboratory (CREL) as well as the Israel National Police (INP) proved being an important relay of information towards a broader range of end-users. Also partner ARTTIC, in charge of dissemination activities, made regular contact with its end-user data base built along SNIFFER in order to create and maintain awareness of the SNIFFER Project and in view of optimised attendance of the final SNIFFER public event. These actions were successful as proved by ad hoc solicitations from stakeholders in the field to join the SNIFFER UAG.\nThe outcome of this regular feedback is reflected by the orientation taken by the SNIFFER version 2 developments (e.g. selectivity of the sensors). It is also reflected by the extensive lab trials ran during the second half of the project, and the final field trials held at Athens International Airport at project end. Concerning this final event, the involvement of UAG member AIA, namely Mr Nikas Anastasos, is to be highlighted. \nEventually, the SNIFFER final public workshop organised at project end in May 2015 in Paris was a meeting point between the project and the end-users that followed project developments as well as others interested in the SNIFFER technology. The main outcomes, the informal exchanges, followed the SNIFFER results presentations (questions asked to the presenters but also informal discussions). Each SNIFFER partner had the opportunity to discuss its specific part with the concerned end-user during the breaks and poster sessions inserted for this purpose in the programme between presentations. The two panel discussions also targeted the positive confrontation between research and end-users, encouraging feedback, comments, suggestion and debate. This closing event was also the opportunity to discuss future perspectives for upcoming research projects. Indeed, the closing panel discussions aimed at considering broader perspectives with regards to artificial sniffing and its applications in the security domain: both the research and the end-user aspects were presented. \n\nTask 8.4 SNIFFER User Workshops\nThe SNIFFER Project organised two public workshops at early year 3 of  the project and at its end. They were both successful. The first public event focused on the rationale of such a SNIFFER R\u0026D Project, from the dog handler to the research point of view; whereas the second event \u2013 presenting final results of the project \u2013 targeted the end-users and \u201cclient\u201d of the SNIFFER Project. Specific reports for each event have been written and cover their organisation and outcome in detail.\nSignificant results: \n\u2022\tDissemination plan successfully implemented\n\u2022\tRegular feedback and input from UAG (considered in its broad understanding): all UCs were covered by the collected feedback\n\u2022\t2 Public Workshops organised\n\u2022\tExploitation defined and promising for a strong SNIFFER impact\nDeviation\nDescription of problem \/ deviation\tSuggested \/ undertaken corrective action\tImpact on other WPs, budget\/PM plan or work plan\nPublic Workshop postponed (waiting for concrete results \u2013 field trials and integrated V1 prototype)\tPostponed event, organisation to be started now to reach a broader audience\tNone\nOverconsumption of human resources at ARTTIC \t\tPartner internal shift from \u201cother direct costs\u201d envelop to \u201cpersonal costs\u201d\tNone\n\n3.9\tWP 9 Project Management\nWP number\t9\tStart date or starting event:\tM1\nWP title\tProject Management \nActivity type\tMGT\nParticipant ID\t1\t14\t\nPart. short name\tCEA\tARTTIC\t\nPM total (for 60M)\t1\t17\t\nPM spent in period 1\t0,56\t8,15\t\nPM spent in period 2\t0,44\t12,70\t\nPM spent total\t1\t20,85\t\nOverview of the progress of the work during the period\t\nARTTIC, the Project Office, and CEA, the Project Coordinator, work closely together and shared project management tasks and responsibilities.\nAt project launch, the Project Office put into place web-based tools for partners to exchange information, be informed of what is expected from the consortium, and be continually updated on project progress. The collaboration between partners was excellent and partners were very involved and active in preparing scientific deliverables.\nSuitable management bodies and procedures for the project were set-up. The General Assembly (GA) has met six times as planned, and the Executive Board (EB) held a teleconference mostly every month. All meeting minutes and deliverables were delivered on time except for a few justified delays. Contractual documents have been agreed upon and signed by all partners, and shared with the EC when needed.\nWork progress in each task \n9.1 Strategic management and financial administration \nAll contractual documents were established, signed and shared with all project partners and the EC if needed during the first months of the project. \nThe pre-financing and interim-payment shares were distributed to all partners according to the agreed timeline and breakdown and in line with the Grant and the Consortium Agreements rules. \nIn terms of strategic management, at project start, the SNIFFER consortium set up its GA, which met twice a year for a consortium meeting, and its EB, which met every month for a project management follow-up teleconference. SNIFFER had all tools needed for success however faced some issues along its course leading to necessary evolution of the work plan and partners\u2019 responsibilities:\n\u2022\tAmendment of the DoW following the midterm review: taking the most of the allocated time and budget through a focused scope and realistic implementation\n\u2022\tPartner CSSC\u2019s voluntary liquidation at the start of project year 3\n\u2022\tA major technical issue: Partner TTS\u2019s risk of being defaulting party and not delivering UC6 sampling device\n\u2022\tPartner EADS\u2019s change of strategy resulting in less integration responsibilities for SNIFFER\n\u2022\tNumerous difficulties leading to a 4-month project extension\nThanks to the efficiency of project management process, these difficulties were identified early enough to be dealt with in the appropriate manner and timing. They were all overcome and eventually did not harm the project success at its end, however leading to quite time and budget consuming activities.\n\n9.2 Project Management Office \nProject management guidelines were agreed upon and implemented at project start. They defined action items for each objective to be dealt with on a daily basis or as necessary. These action items were adjusted as needed during the project to match the project requirements and learn from past lessons.\nBesides the initial set up of the project management process, methods and tools, regular milestones were to be met during the project to assess the efficient project management. The calendar as well as the scope of the project were slightly redefined at project midterm in order to better fit to what was realistically possible, considering the status of the SNIFFER developments at this stage (e.g. specific focus on 2 usage cases), and not to lose time with little added-value actions (e.g. replacing midterm field trials by extensive lab trials). Regular feedback was put in place leading to regular updates of the implementation plan and the corresponding action lists for both the managing team and the partners. \nBy doing so, calendar, quality and scope were respected despite difficulties such as the voluntary liquidation of a partner or the change of strategy of another one. Due to internal restructuring and challenging engineering feats for integration, the consortium and European Commission Project Officer agreed upon a 4-month extension to the project. \nEventually, as the SNIFFER project came to its end with successful technical results, promising exploitation and fruitful further collaboration in between partners, it is to be noted that efficient project management was successfully achieved during the project duration. \nSignificant results: \nThe most significant result of WP9 is demonstrated in the fact that the collaboration between partners was very good, and sometimes even excellent.\nThe SNIFFER project management was carried out on a day-to-day basis, relying on web-based tools available at all times to all partners. The Project Management guidelines including a management plan were established at project start and agreed by all partners (D921). The management bodies were elected and meet on a regular basis. More precisely, to monitor the project, a risk register has been established and is regularly updated (first iteration shared with the EC through D922). Internal reporting is monthly done.\nDeviations\nDescription of problem \/ deviation\tSuggested \/ undertaken corrective action\tImpact on other WPs, budget\/PM plan or work plan\nTechnical results needing a re-focus of project scope for efficient use of allocated time and budget\tDiscussions at consortium level, GA and then during the mid-term review meeting with EC PO and reviewers  \tSee DoW Amendment #1\nCSSC voluntary liquidation\tReconsideration of ethical aspects, also with regards to scope refocus\tExtra budget for technical work and field trials\nPartner TTS\u2019s risk of being defaulting party and not delivering UC6 sampling device\tIntense project management, delayed interim payment upon evidence of work, web demonstration of the developed device working + internal testing at partner INP\n\tPotential failure of UC6\nAIRBUS (Former EADS) change of strategy\tInvolvement of a new laboratory of CEA in the project: CEA-DACLE + further involvement of UNIMAN with regards to integration work\tDelay in the technical developments and integration work + Impact on future exploitation\nDelay in technical results\tProject extension of 4 months\tFailure to provide final assessment of the fully integrated devices\nPotential Impact:\n3.10\tPotential impact (including the socio-economic impact and the wider societal implications of the project)\n\nMany sensitive controls are carried out at border checks by trained dogs, which cannot be neither replaced, nor be expected to operate on a 24\/7 basis. SNIFFER devices cover the variety of border security situations (contexts and targets) where dogs are used today. The devices, once developed as market products, will be able to operate on a 24 hours\/7 days per week basis and allow dogs to be used in situations where they can potentially make a difference. \nSNIFFER, in particular with the V2 technology developed during the 2nd period of the project, proved being able to go farther than dogs as the sensitivity of SNIFFER devices will keep increasing. Dogs are also unable to push air onto objects to collect particles or gas molecules as is intended for the automatic artificial sniffer portal proposed by SNIFFER to control suitcases on luggage belts. Through these achievements, border security would be significantly enhanced as the difficulty for illegal traffics of all kinds (drugs, tobacco, illegal immigration...) to cross border lines, especially in airports, would dramatically increase. At the same time, border security would be improved as regards to possible terrorist attacks. Although SNIFFER focused on specific usage cases leaving aside container checking and human odour detection, the project results are quite promising with regards to these potential exploitation fields. It is to be noted that the SNIFFER technology will be further developed through the H2020 Project C-BORD where some of the SNIFFER submodules will be adapted and tested for the detection of illicit substances in containers. In general, the expected impact of SNIFFER will contribute to the overall field of security. Many security applications would benefit from the high-sensitivity low-cost technology provided by SNIFFER (in comparison with COTS), for example: the search of casualties after a disaster, the detection of dangerous gases close to an industrial facility or the search of drugs by police forces. Indeed the SNIFFER instrument developed is much promising and a mature result. As a sensing electronic submodule capable of sensing substances it can be exploited in many markets such as that of the access control in every critical infrastructure and use of the equipment from law enforcement agencies.\n\nFurthermore SNIFFER will contribute to many other application fields such as: health (breath analysis, other body odours, faeces), food safety (e.g., to detect problems with recycled bottles) and environmental monitoring (e.g., air and water quality). SNIFFER results obtained during the process development of OBPs may be helpful in the case of development of other OBPs for other applications such as pharmaceuticals, diagnostic, environment, biotechnology and nanotechnologies, industrial monitoring. Moreover, the development of MUP targeted to different analytes can lead to the development of sensors based on different technologies, which may be used in various fields (pharmaceuticals, environmental). Also, the innovative characteristics of the SNIFFER work sits in the durability and re-usability of the sensors which has shown the potential for evolution of the system for other applications. Another field of application will be the detection of volatile organic compounds for medical diagnostic and air\/water monitoring. At industrial level, broader potential and outcome from the OBP-sensing- and ESP-sampling developments can be envisaged. The technology by its own could also be used in testing the air quality as well as to test health and safety issues in structure and manufacturing processes for instance. \n\n\n\nSpecific individual plans for exploitation of project results are summarized below.\nMajor avenues for commercial exploitation include the following list of SNIFFER enabled products:\n\u2022\tThe microcantilever based Multisensor chemical detection module (CEA, UNIMAN, ESIEE) addresses the market of multiple sensor applications (potential of 500k\u20ac in 5 years). As it reached TRL 5 at the end of the SNIFFER Project, time to market is estimated to 3 years. Competitive assets sit in the complementary and concurrent design of both the sensors, the support electronic and the odour detection algorithm.\n\u2022\tThe diamond MEMs with Polysilicon gauges (ESIEE) address the market of integrated sensor industrial for specific application. It involves innovative and complementary process between partners CEA and ESIEE. The potential is evaluated to be 2M\u20ac in 3 year-time from now. \n\u2022\tThe recombinant proteins (UNIMAN) reach pharmaceutical, environmental, security sectors for which the size of market is still to be defined. Potential has however been estimated to 1M\u20ac in 3 years, up to 5M\u20ac in 5-10 years. Current competitive assets are the novel, stable biorecognition elements as well as SNIFFER partner available facilities for production.\n\u2022\tThe SNIFFER Instrument (UNIMAN) could reach pharmaceutical, environmental and security markets in 3 year-time. Potential has been estimated to 1M\u20ac in 3 years, 2M\u20ac in 5 years and up to 5M\u20ac in 10 years.\n\u2022\tThe pre-concentrator components (ARMINES, EPFL) address the market of small detection devices for specific applications (environment, security). The main barrier to market sits in getting the preconcentrators to be manufactured. A major competitive asset is the acquired expertise on the manufacturing and the user of pre-concentrator components.\n\u2022\tWith regards to OBPS, Ligand Binding Proteins (GTP) will address the pharmaceuticals, diagnostic, environment, biotechnology and nanotechnologies and industrial monitoring market in 2 to 3 years. Potential is estimated to 1M\u20ac in 5 years. A major asset is the new protein biosensors. Major Urinary Proteins (UNIPD) will address the pharmaceutical, environmental and security markets, also possibly generating 1M\u20ac in 5 years. The major asset is the custom-tailored proteins for biosensors.\n\u2022\tThe ESP-sampling and thermos-desorption system (Airbus Group Innovations) will reach new fields of application beyond the security field through the detection of failure or safety issues with regards to carbon fibre manufacturing processes.\n\u2022\tThe SNIFFER Monitoring portal (TTS) reaches the major transportation hubs, shopping centres, government buildings, sports and other entertainment arenas markets. Its integration into 4 TTS products \u2013 ShoeSafe, CompactSafe, CarrySafe, and CargoSafe \u2013 would generate revenues from $375 000 in 3 years for ShoeSafe to $40 000 000 in 10 years for CarrySafe and CargoSafe. There are almost no competitors with the same level of technology for particle extraction. With the use of an activated carbon filter, it is also possible to extract vapors from the inspected baggage.\n\nThere are also SNIFFER non-technical results developed with intention to exploit following the project, in order to optimise the SNIFFER impact. In particular, this includes the potential for partner 3DSA to take on a role as consultant for SNIFFER users following the project. This will include consulting and training services, as well as optimisation of security protocols and policies for SNIFFER devices.  It also includes partner ARTTIC activity, as involved in other FP7 and H2020 projects as part of its core business, which intends to keep up the work carried out during SNIFFER dissemination activities. The objective will be for the community built around SNIFFER (end-users, research, institutions interested in artificial sniffing and\/or detection for border security) to benefit from its network in future projects \/ events in the field.\nWith regards to patent, patent on \u201cMicrocantilever based Multisensor chemical detection module\u201d has been filed during the project between CEA\/ESIEE-Paris\/UNIMAN. This could be exploited through licensing to a third party of through the creation of a start-up company supported at least by CEA depending on the results from market study.\nPartner UNIMAN also envisages to develop new biosensors based on OBPs, build further EU project proposals in the security area based on SNIFFER developments to take TRL levels to TRL 6 and above and publish scientific papers with partners CEA, UNIPD on new developments on OBPs and MUPs for chemical sensing. Partner EPFL is approaching companies developing portable gas analyser systems for technology transfer of the gas preconcentrator developed on slicon and envisages the development of polymeric based disposable gas detection systems based on printing technology.\nEventually, the technology developed by CEA and other partners within SNIFFER will be further developed through the H2020 project C-Bord.  \nList of Websites:\nwww.sniffer-project.eu","sourceUpdateDate":"2015-12-16 00:00:00","contentCreationDate":"2019-05-22 17:20:46","contentUpdateDate":"2019-05-22 17:20:08","lastUpdateDate":"2024-02-06 08:57:43","archivedDate":"2024-06-23 00:00:00","relations":{"associations":[{"availableLanguages":"de,en,es,fr,it,pl","rcn":"170344","id":"170344-innovative-odour-detection-technology-to-secure-borders","title":"Innovative odour detection technology to secure borders","teaser":"Dogs and their superior odour detection capabilities are an invaluable asset to border services. An EU initiative is designing artificial sniffer technology to complement and leverage the work done by dogs.","contentUpdateDate":"2015-12-29 15:59:21","archivedDate":"2024-06-10 00:00:00","relations":{"categories":[{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"indust","title":"Industrial Technologies","displayCode":"\/Industrial Technologies","attributes":{"classification":"applicationDomain","type":"isApplicableIn"}},{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"brief","title":"Result in Brief","description":"Article: Classic and enhanced","displayCode":"\/Result in Brief","attributes":{"classification":"collection","type":"belongsTo"}},{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"editorial","title":"CORDIS Editorial","description":"Written by CORDIS","displayCode":"\/CORDIS Editorial","attributes":{"classification":"source","type":"isProvidedBy"}}],"associations":[{"language":"en","availableLanguages":"de,en,es,fr,it,pl,any","uri":"\/docs\/results\/images\/2015-12\/163709.jpg","mimetype":"image\/jpeg","size":"25476","copyright":"\u00a9 Anne Mette Lykke","contenttype":"webItem","attributes":{"represents":"article","source":"editorial","type":"relatedImage"}}]},"contenttype":"article","attributes":{"source":"editorial","type":"relatedArticle"}},{"availableLanguages":"en","rcn":"1958119","legalName":"COMMISSARIAT A L ENERGIE ATOMIQUE ET AUX ENERGIES ALTERNATIVES","address":{"country":"FR"},"contenttype":"organization","attributes":{"source":"sesam","type":"relatedOrganization"}},{"availableLanguages":"en","rcn":"102348","id":"285203","acronym":"SNIFFER","title":"A bio-mimicry enabled artificial sniffer","archivedDate":"2024-06-18 00:00:00","relations":{"categories":[{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"project","title":"Project","description":"Project factsheets","displayCode":"\/Project","attributes":{"classification":"collection","type":"belongsTo"}},{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"\/25\/73\/453\/58525161\/66357197","title":"biosensors","displayCode":"\/engineering and technology\/electrical engineering, electronic engineering, information engineering\/electronic engineering\/sensors\/biosensors","attributes":{"classification":"euroSciVoc","type":"isInFieldOfScience"}},{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"\/23\/47\/307","title":"software","displayCode":"\/natural sciences\/computer and information sciences\/software","attributes":{"classification":"euroSciVoc","type":"isInFieldOfScience"}},{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"\/27\/83\/501","title":"domestic animals","displayCode":"\/agricultural sciences\/animal and dairy science\/domestic animals","attributes":{"classification":"euroSciVoc","type":"isInFieldOfScience"}},{"language":"en","availableLanguages":"en","code":"CP-FP","title":"Small or medium-scale focused research project","displayCode":"\/Small or medium-scale focused research project","attributes":{"classification":"projectFundingSchemeCategory","type":"relatedProjectFundingSchemeCategory"}},{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"corda","title":"CORDA","description":"RTD research data warehouse","displayCode":"\/CORDA","attributes":{"classification":"source","type":"isProvidedBy"}}],"associations":[{"availableLanguages":"de,en,es,fr,it,pl","rcn":"861","id":"FP7-SECURITY","code":"FP7-SECURITY","frameworkProgramme":"FP7","pga":"FP7-SECURITY","title":"Specific Programme \u0022Cooperation\u0022: Security","shortTitle":"Security","parent":{"rcn":"846\/1401\/837","code":"FP7-COOPERATION\/EC\/FP7"},"archivedDate":"2024-06-12 00:00:00","relations":{"categories":[{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"legalbasis","title":"Legal Basis","description":"Programme: Legal basis for research funding","displayCode":"\/Legal Basis","attributes":{"classification":"collection","type":"belongsTo"}},{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"editorial","title":"CORDIS Editorial","description":"Written by CORDIS","displayCode":"\/CORDIS Editorial","attributes":{"classification":"source","type":"isProvidedBy"}}]},"contenttype":"programme","attributes":{"source":"corda","type":"relatedLegalBasis"}},{"availableLanguages":"en","rcn":"17134","id":"FP7_SEC-2011.3.4-2","code":"SEC-2011.3.4-2","frameworkProgramme":"FP7","title":"\u201cArtificial sniffer\u201d- Capability Project","parent":{"rcn":"11631\/10351\/11558\/861\/846\/1401\/837","code":"SEC-3.4\/SEC-3\/SEC\/FP7-SECURITY\/FP7-COOPERATION\/EC\/FP7"},"archivedDate":"2024-06-14 00:00:00","relations":{"categories":[{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"topic","title":"Topic","description":"Programme: Part of the Commission work programme implementing legal basis","displayCode":"\/Topic","attributes":{"classification":"collection","type":"belongsTo"}},{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"editorial","title":"CORDIS Editorial","description":"Written by CORDIS","displayCode":"\/CORDIS Editorial","attributes":{"classification":"source","type":"isProvidedBy"}}]},"contenttype":"programme","attributes":{"source":"corda","type":"relatedTopic"}}]},"contenttype":"project","attributes":{"source":"sesam","type":"relatedProject"}},{"language":"en","availableLanguages":"en,any","uri":"\/docs\/results\/285\/285203\/final1-2015-07-30-sniffer-rep-final-report-overall-project-vf.pdf","mimetype":"application\/pdf","size":"5124831","contenttype":"webItem","attributes":{"source":"sesam","type":"relatedResultAttachment"}}],"categories":[{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"pubsum","title":"Report Summary","description":"Result: Publishable summary (FP7 SESAM) and Summary for publication (H2020 SyGMa)","displayCode":"\/Report Summary","attributes":{"classification":"collection","type":"belongsTo"}},{"language":"en","availableLanguages":"de,en,es,fr,it,pl","code":"sesam","title":"SESAM","description":"RTD FP7 reporting system (including FP6 Marie Curie)","displayCode":"\/SESAM","attributes":{"classification":"source","type":"isProvidedBy"}}],"regions":[{"name":"France","rcn":"258902568","nutsCode":"FR","euCode":"FR","isoCode":"FR","attributes":{"source":"sesam","type":"relatedRegion"}}]},"modelVersion":"132","icaVersion":"20.0.0-48f81e07"}