Service Communautaire d'Information sur la Recherche et le Développement - CORDIS


BIOGRAD Résumé de rapport

Project ID: G5RD-CT-2000-00354
Financé au titre de: FP5-GROWTH
Pays: Italy

Laboratory procedure for wear testing of Al2O3 hip prosthesis on a biaxial rocking motion hip simulator

EC-JRC-IHCP performed wear test of Biolox forte(TM) hip components using a Biaxial Rocking Motion (BRM) hip joint simulator. The results obtained, following the ISO 14242 standard, led to the definition of the provisions that has to be taken against the operational sources of error.

Reported by

Institute for Health and Consumer Protection, European Commission - Joint Research Centre
TP 500; Via Enrico Fermi 1
21020 Ispra - VA
See on map