Final Report Summary - SCANCER (Specialized cantilevers for cancer research)
The measurement of cellular or tissue stiffness consists in essence of two parts: a force that is applied to the cell/tissue and the deformation that such force causes is measured. By relating these two parameters, one can estimate the stiffness. AFM is one of the tools that are typically used to perform these measurements in a very reliable and accurate fashion. Nevertheless, this technique has important drawbacks in terms of complexity, fragility and difficulty together with a very low throughput.
This project, SCANCER (Specialized cantilevers for cancer research), has aimed to ease AFM spectroscopy technique on single cells by developing specialized SU-8 probes. SU-8 has a low Young’s modulus achieving lower spring constants to the levers with thicker structures for better stability in liquid; the tip geometry is easily tunable to be less sharp, beneficial for a more gentle contact with the cells; and SU-8 provides a cost-effective fabrication.
SCANCER has also explored the capability of the SU-8 probes coated with graphene for fundamental research on graphene interactions with cells and opening the possibility of new functionalization and electrical measurements together with a longer preservation of the tip geometry.
SCANCER also has aimed to increase throughput by using 2-dimensional (2D) probe arrays on a homemade interferometric AFM system. By performing the measurement of multiple cells in parallel, stronger statistics can be obtained in the same amount of time. This work demonstrates that parallel force spectroscopy is a valid approach to significantly make an improvement in the field of cancer research by means of faster acquisition of elasticity data.