Since the End of Reporting Period 1 (28th of February 2022) until project completion (29th of February 2024), MODE has achieved the following key results:
• Updated patent and IPR strategy.
• Updated risk matrix and mitigation actions.
• Developed and verified new sensor-patch with improved performance in terms of wearability, water, resistance, stability, and bioimpedance measurement accuracy.
• Developed and verified production testers responsible for automatic programming, verification of performance, etc. suited high-volume manufacturing.
• Improved the on-device hydration calculation algorithm by implementing quality assurance, the proprietary tissue model, other compensation filters using data from the other sensors on the sensor-patch, hereunder accelerometer and thermistor.
• Developed prototype for IoT-gateway and API Bluetooth interface, incl. integrator manual.
• Finalized clinical validation report from study on nursing home residents.
• Performed all safety and performance testing required for regulatory submission and clearance, incl. biological safety testing, electrical safety, bench-performance testing, etc.
• Completed technical files for 510(K) Clearance and CE-marking, incl. verification & validation documentation, labelling, instructions for use, etc.
• Implemented a full QMS according ISO13485 and performed internal audits and notified body pre-assessments proving compliance with the standard.
• Updated business plan, competitive analysis, and financial forecast.
• Secured three commercial co-development agreements with multinational commercial partners in med-tech, military, welfare technology markets.