Objectif
The oxygen content of silicon is an important parameter for the manufacturing of integrated circuits and other electronic devices. The measurements are based on infrared absorption spectrometry. Results of infrared absorption measurement often differ by several percent between different laboratories. Industry has therefore requested the BCR programme to produce silicon reference materials (zero specimens needed in the measurements as well as samples to check infrared spectrometers).
RESULTS
Intercomparisons have been organized to improve the measurements of the actual oxygen content and of the infrared absorption.
Two reference materials have been certified.
CRM 368: silicon with very low 0 content, to be used as "zero"
specimens: 69 +/- 19 ng/g
Certification report: EUR 12927 (1990)
CRM 369: silicon with 0 contents in the range of industrial interest, for testing the accuracy (photometry + software) or the IR Instruments:
absorption coefficients in the range 2.2 - 3.8 cm{-1}
oxygen contents in the range 8 - 13 ug/g
Certification report : EUR 12928 (1991)
Champ scientifique
Thème(s)
Data not availableAppel à propositions
Data not availableRégime de financement
CSC - Cost-sharing contractsCoordinateur
OX11 0RA Didcot
Royaume-Uni
Voir sur la carte
Participants (16)
Orleans
Voir sur la carte
28100 Novara
Voir sur la carte
PARIS
Voir sur la carte
12205 BERLIN
Voir sur la carte
SW7 2AZ London
Voir sur la carte
Gent
Voir sur la carte
60486 Frankfurt am Main
Voir sur la carte
Geel
Voir sur la carte
Gent
Voir sur la carte
44139 Dortmund
Voir sur la carte
SO Southampton
Voir sur la carte
Great Malvern
Voir sur la carte
81739 München
Voir sur la carte
74072 Heilbronn
Voir sur la carte
RG6 6AF READING / SILCHESTER
Voir sur la carte
84489 Burghausen
Voir sur la carte